Owner's manual

www.amico.com 27
If necessary, the line regulator replacement can be performed while manifold is in service. However, this should only
be done by qualied technicians experienced in servicing medical equipment.
1. Close the two ball valves on inlet and outlet of line regulator to be replaced.
2. Loosen union nuts on the two closed isolation valves.
3. Vent pressure from the bank, which was shut o in step 2 by depressing the bleed valve on side of the regulator.
4. Ensure the orientation of ttings (direction of gas ow) is the same on the replacement regulator. Install ttings on
new regulator.
5. Inspect O-ring removed from the ball valve union. If damaged, replace O-ring.
6. Verify the correct orientation of ttings on the regulator as some slight adjustment may be required to align
with the ball valves without causing stress to the ttings.
7. Hand-tighten the two ball valve union nuts and then with a wrench, snug lightly to ensure the ttings are secure.
NOTE: Do not over tighten unions, as it is an o-ring seal.
8. Open the ball valves and set the line regulators output pressure as described in Line Regulator Pressure Adjustment
procedure.
Line Regulator Replacement