6.4 Medical Air Systems Glossary
Denitions
ICFM • Conditionsattheinletofthecompressorpriortoanyrestrictionsor
temperature changes. (i.e., inlet lters, inter-coolers, discharge manifolds
etc).
Displacement • Theoreticalphysicalvolumeoftheairinthecompressorchamberbased
upon100%pumpingeciency,withnoallowancesmadeforheat,
friction, clearances, or other losses in the compression cycle.
PeakCalculated • Themaximumestimateddemandamedicalfacilitywillrequireofa
Load (PCL) medical air system.
• CalculatedatSCFMat100psig.
Oil-FreeCompressor • Compressorwithoilwithinthecompressorbutabsentfromthe
compression chamber by means of seals and visible vented distance
piece, and a dry compression chamber.
SimultaneousDemand• Operatingreferencetoaconditionwhereallcompressors(LeadandLag
compressors) run simultaneously to satisfy demand in excess of lead
compressor(s) capability.
DewPoint • Temperatureandpressureatwhichmedicalairwillcondensewater
vapor into liquid water within the medical air pipeline system.
ContinuousDuty• Aircompressorswhichcanoperatecontinuously(24hoursperday)if
Rated necessary but normally only operate on demand.
LPM • AbbreviationforLitersperminute;ameasureoftheowrateofagas.
QuadruplexSystem • Medicalairsystemwithfourcompressors(4),eachsizedfor50%PCL.
ACFM • ActualCubicFeetperMinuteisanexpressionofactualairvolume,
generally corrected for and reference to a particular pressure.
ContinuousDuty • Operationalreferencetocompressorsoperating24hoursaday,
continuously.
DuplexSystem • Systemscomprisedoftwo(2)compressors,eachratedfor100%Peak
Calculated Load (PCL).
DesiccantDryer • Consistsoftwotowers,wherebythedesiccantbedinonetowerdriesthe
air stream while a purge of dry air regenerates the desiccant bed in the
other tower.
NTP • “NormalTemperatureandPressure,generallyacceptedas20˚C(70˚F),
14.969psi/29.92”Hgbarometricpressure,and36%relativehumidity.
DewPoint/COMonitor • Monitorsdewpointandcarbonmonoxidelevelsinmedicalair.
Oil-LessCompressor • Compressorwithnooilcontainedwithinthecompressor.
SCFM • StandardCubicFeetPerMinuteisanexpressionofairatNTP.
TriplexSystems • Medicalairsystemwiththreecompressors(3),eachsizedfor100%PCL.
7