User guide

Mounting Information
Approval Signature Date Phone No.
NOTE: Riser plate to be installed in ceiling with appropriate mounting structure and anti-sway sup-
port. The methods of support shown are provided for information only.
NOTE:
1. All labour and material for installation of Riser to Mounting Structure supplied by others.
2. General Contractor has  nal responsibility for the strength and stability of the mounting structure that is
supplied by others.
3. Riser plate to be mounted 4.5” (114.3mm) above  nished ceiling.
4. Mounting of column unit to riser plate by (4) 1/2” threaded rods, provided by Amico.
5. Space between top of Riser Plate and structural ceiling must be minimum 10”(254mm).
6. Riser Pipe(s) extend 8” (203.2mm) above riser plate.
7. Pre-ship of the riser assembly is available upon customers request.
STRUCTURAL CEILING
17” Sq.
[431.8mm Sq.]
CEILING COVER
(FIELD MOUNTED)
FINISHED CEILING
COLUMN
MOUNTING
PLATE
1/2” [12.7mm] DIA. MOUNTING BOLTS FUR-
NISHED WITH RUOGH-IN PLATE. TYP. OF 4, See
note 1.
MIN. 13” [330.2mm] MAX.
16” SQ CEIILNG CUT-OUT
FOR COLUMN
17” Sq.
[431.8mm Sq.]
14.75” Sq.
[374.65mm Sq.]
9.5” Sq.
[241.3mm Sq.]
RISER
PLATE
ANTI-SWAY SUPPORT STRUCTURE
(SUPPLIED BY OTHERS)
1.00” [25.4mm]
4.5”
[114.3mm]
3/4” [19.5mm] or 7/8” [22.23mm]
THREADED ROD (X4) (SUPPLIED BY OTHERS)
12 Amico Pipeline