Amico recommended circuit schedule based on layout
Custom circuit schedule. Please specify on circuit chart
Receptacle Type:
US/CAN
*BRITISH (simplex only)
*GERMAN (simplex only)
Circuit Chart
Electrical Diagram (Bottom View)
Approval Signature Date Phone No.
NOTE:
1. Two electrical/communication/medical gas devices
cannot be placed together at any corner.
2. If type of wire & gauge is not specified, Amico
will use the following:
For US/International
LIVE/HOT - 12 Ga. THHN Wire Black
NEUTRAL - 12 Ga. THHN Wire White
GROUND - 10 Ga. THHN Wire Green
For Canada
LIVE/HOT - 12 Ga. RW90 Wire Black
NEUTRAL - 12 Ga. RW90 Wire White
GROUND - 10 Ga. RW90 Wire Green
3. Conduit: 9/16” [14.28 mm]  exible metal conduit.
4. All conduits and wires will extend a minimum of 12”
[304.8 mm] above  nish column for connection
to riser plate.
5. A pull-string will be provided to be connected by
others for all ground jacks and data provision.
6. British or German Receptacle cannot be positioned
side by side i.e.: A1 and A2.
*Please specify electrical details:
Type of wire:
THHN XHHW RW90
AWG Wire Gauge: Wire Color:
10 & 12 LIVE/HOT
10 & 12 NEUTRAL
10 GROUND
Isolated Power
Please select one of the following
All receptacles shall be mounted with
ground pin “DOWN”
All receptacles shall be mounted with
ground pin “UP (Default)