Owner's manual

24 Amico Pipeline
Appendix F
WIRING DIAGRAM: LCD DISPLAY MODULE - MASTER MODULE
#22 Gauge twisted
pair shielded cable
6"-8” [0.1 m-0.2 m]
supplied
#22 Gauge twisted
pair shielded cable
5,000' [1,500m] max.
Sensor
Module
DISS
Junction Box
(by others)
Remote location
Alarm
Master
Module
Note: Jumper any unused points
on the Master module.
U 2
U 6 U 5
U4
U4
U3
NC
C
NC
C
NC
C
NC
C
NC
C
NC
C
NC
C
NC
C
Source
Equipment
C H 1 C H 2 C H 3 C H 4 C H 5 C H 6 C H 7 C H 8 C H 9 C H 1 0
ON
1
2
3
4
OXYGEN
ON
1
2
3
4
5
6
7
8
9
10
+++++++
+
++
#22 Gauge twisted
pair shielded cable
100’ [30.5 m] max.
AC Supply
115 to 220 VAC
DC Power Cable:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle Switch
B4 B3 B2 B1
Source
Equipment
Red
Black
SETUP
CHANGE UP
CHANGE DOWN
SELECT/MUTE
NOTE:
Jumper any unused points on
the Master module
Turn OFF dip-switches for any
unused points (Location SW-2)
S1 S2
To Amico: Master
Module or Building
Management System