Operating & Maintenance Manual Alert-3 LCD Alarm v2.0 1ST FLOOR ROOM 12 2ND FLOOR ROOM 25 4TH FLOOR ROOM 47 3RD FLOOR ROOM 37 MAINTEN EXT 302 5TH FLOOR ROOM 58 w w w . a m i c o .
Contents User Responsibility Introduction Features 3-4 Description of the Alarm Shipment Details The Alarm Back Box The Frame/Module Assembly System Power Supply LCD Module Sensor Module 5-6 5 5 5 6 6 6 Installation Guide The Alarm Back Box For Local Sensor Only Standing Pressure Test Sensor Local Remote Frame/Module Assembly System Power Supply Sensor Module Gas Display Local Remote LCD Display Module LCD Display Setup Closing The Frame/Module Assembly 7 - 11 6 6 6 6-7 6 7 7 8 9 - 10 9 9 9 9 10 10 - 1
Contents Troubleshooting 17 - 19 Appendix A Wiring Diagram: LCD Motherboard 20 Appendix B Wiring Diagram: Auto-Switch Power Supply 21 Appendix C Wiring Diagram: LCD Display Module - Alarm Buzzer 22 Appendix D Wiring Diagram: LCD Display Module - Local Sensor 23 Appendix E Wiring Diagram: LCD Display Module - Remote Sensor 24 Appendix F Wiring Diagram: LCD Display Module - Master Module 25 Appendix G Technical Specifications 26 www.amico.
User Responsibility Amico Microprocessor Based Alarm USER RESPONSIBILITY The information contained in this Installation, Operation and Maintenance Manual pertains only to the Alert-3 microprocessor based digital LCD Alarm. This product will perform as described in this manual when assembled, The information contained in this Installation and Operation Maintenance operated, maintained and in accordancebased with digital the installation instructions provided.
Features • Microprocessor based digital LCD and individual microprocessor on each sensor module • Gas-specific sensors can be mounted locally or remotely, up to 100 feet [30.5 m], utilizing 22 gauge shielded twisted pair (2 shielded wires) • DISS gas-specific sensor housed in a tamper-proof enclosure.
Description of Modules The Alert-3 LCD Alarm is a high technology microprocessor based module: COMMON TO ALL ALARMS SYSTEM POWER SUPPLY The System Power Supply has been pre-installed into the back box assembly. The System Power Supply converts the AC voltage supply to the alarm into two voltages: 5 VDC (regulated) required by the microprocessor hardware and 15 VDC (unregulated) required by the buzzer and the LCD.
Installation Guide STEP 1: THE ALARM BOX Install the back-box to the studs of the wall at the desired height. Ensure that the box is securely in place. The mounting brackets are adjustable to suit the thickness of the wall. MAKE SURE the box is parallel, squared and flush with the finished wall surface to ensure that the frame assembly will fit properly. STEP 2: FOR LOCAL SENSOR ONLY If the sensors are to be mounted locally (inside the back box), the pipe stubs must be connected to the pipeline.
Installation Guide B: REMOTE (outside the back box) Push the sensor module hex-nut and nipple adapter up into the demand checkvalve. With a wrench, tighten the nut so that it makes a good seal. NOTE: Pressure on sensors are not to exceed 250psi for Pressure sensors and 30” for Vacuum sensors REMOTE (Outside the Back Box) 1. Connect a Tee (supplied by others) to the pipeline with a 1/4" NPT female with a 1/4” NPT female connection that will accept A m ii.
all warranty on the alarm. Statements in this manual preceded by the words WARNING, CAUTION, DANGER and NOTE are of special significance. Please read these sections carefully. Installation Guide STEP 6: SYSTEM POWER SUPPLY TURN OFF POWER SWITCH before changing any modules and/or disconnecting any cables. Failure to do soWARNING: can cause the fuse to blow, damaging the circuitry. denotes steps which can prevent injury. i. Ensure that the ON/OFF switch is in the OFF position. ii.
Installation Guide STEP 7: SENSOR MODULE B: LOCAL (inside the back box) i. The sensor module is provided with a 6”-8” [0.1 m - 0.2 m] twisted pair of wires. One wire is red (positive) and the other wire is black (negative). Connect the wires to the display module as shown in Appendix D. Take the red wire from the sensor and attach it to terminal “Sensor +” on the display module. Take the black wire from the sensor and attach it to terminal “Sensor -”.
Installation Guide STEP 9: LCD DISPLAY SETUP i. Press the SETUP button (B1) and press the SELECT button (B4) ii. Volume control: 90, 80, 70, 60 – press CHANGE UP/DOWN to change noise level iii. Press SELECT for LCD brightness and press CHANGE UP/DOWN to change LCD brightness iv. Press the SELECT button (B4) * The following should be displayed.
Programming Gas Locations 1. Use the Notepad program to enter information for gas location. Each line can hold up to a maximum of 16 characters. 2. Two lines may be used per individual gas. 3. The order of the text must go in order of gas; meaning the first two lines of text shall represent the 1st gas location, the next two lines of text shall represent the 2nd gas location, and so forth.
Model Numbers LCD ALARM A3AR-L-XXXXXXXX R = Remote/Local “L” Defines the Language: ”X” Defines the Type of Gas: C = Conversion (Retrofit) U = English (NFPA) Oxygen = O E = English (CSA/ISO) Medical Air = A F = French (CSA/ISO) MedVac = V S = Spanish (NFPA/ISO) Nitrous Oxide = 2 Nitrogen = N Carbon Dioxide = C Waste Anesthetic Gas Disposal (NFPA) = W Anesthetic Gas Scavenging System (CSA/ISO) = E Instrument Air = I ALERT-3 SENSOR MODULE A3P-SENS-L-GAS “Gas” D
Spare Part Numbers SENSORS Model Number Description A3P-SENS-E-AIR Sensor Module ISO-AIR Eng. Alert-3 A3P-SENS-E-CO2 Sensor Module ISO-CO2 Eng. Alert-3 A3P-SENS-E-AGS Sensor Module ISO-AGS Eng. Alert-3 A3P-SENS-E-N2O Sensor Module ISO-N2O Eng. Alert-3 A3P-SENS-E-NIT Sensor Module ISO-NIT Eng. Alert-3 A3P-SENS-E-OXY Sensor Module ISO-OXY Eng. Alert-3 A3P-SENS-E-VAC Sensor Module ISO-VAC Eng. Alert-3 A3P-SENS-E-IAR Sensor Module ISO-IAR Eng. Alert-3 A3P-SENS-U-AIR Sensor Module USA-AIR Eng.
Spare Part Numbers DEMAND CHECK VALVES Model Number Description S-DIS-DEMC-AIR DISS Demand Check Valve 1/4” MNPT - AIR S-DIS-DEMC-CO2 DISS Demand Check Valve 1/4” MNPT - CO2 S-DIS-DEMC-NIT DISS Demand Check Valve 1/4” MNPT - NIT S-DIS-DEMC-N2O DISS Demand Check Valve 1/4” MNPT - N2O S-DIS-DEMC-EVA DISS Demand Check Valve 1/4” MNPT - EVA S-DIS-DEMC-OXY DISS Demand Check Valve 1/4” MNPT - OXY S-DIS-DEMC-VAC DISS Demand Check Valve 1/4” MNPT - VAC S-DIS-KIT-OXY DISS Demand Check, Nut and Nippl
Dimensions LCD ALARM Front View Side View Inch [mm] 2 1/10 [53.34] 0.81 [20.57] 1.2 [30.48] 1.2 [30.48] 1.2 [30.48] 1.2 [30.48] 1.2 [30.48] 1.2 [30.48] 1.2 [30.48] Only supplied with Local Sensors 3/8” [9.53mm] (O.D.) 1/4” [6.4mm] (I.D) Type “K” Copper Pipe 7 3/16 [182] NOTE: DISS demand check valves by 1/4” NPT, supplied with each sensor 1 1/4 [32] 6.375 [162] 9 [228.6] 10.25 [260.35] 7.50 [191] Brazing Filler Metal AWS Spec. BAg-7 3.97 [100.84] 12 [304.8] 1/2-14NPSM THD.
Troubleshooting Symptom An error or “LOW ALARM” LCD screen No power on the alarm Power light is ON, however there is no display on LCD screen No audible alarm Audible signal will not silence Cause Corrective Action The Microprocessor detected a fault and has shut down Turn power switch to OFF position. Wait for at least 5 seconds before turning ON the power. The program will reset itself.
Troubleshooting Symptom Cause Gas reading incorrect Display shows “NO SENSORS” Loose connection of DISS fittings Ensure that the sensor module is properly connected to the DISS demand checkvalve Sensor module is not properly wired to the display module Ensure that the sensor module is properly wired to the LCD module by using wiring diagram in Appendix D or Appendix E Defective sensor or requires calibration Replace the sensor module Defective LCD display module Replace the LCD display module No
Appendix A WIRING DIAGRAM: LCD MOTHERBOARD D C C An on Pow nu ne er nc ct C ia to ab to le : rM od ule DC Connector Mute Button (located on reverse side of board) Sensor Terminal Buzzer LCD Cable Connector Reset Button 12V DC Output SETUP CHANGE UP CHANGE DOWN SELECT/MUTE LCD Power B4 B3 B2 B1 Microchip SD Card Four Push Buttons Dry Contacts (to Building Management System/Master) www.amico.
Appendix B WIRING DIAGRAM: AUTO-SWITCH POWER SUPPLY ACACSupply Supply 115toto220 220 VAC VAC 115 G N L L - Live L - Live N - Neutral N - Neutral G - Ground G - Ground Ground Ground Toggle Switch Toggle Switch Fuse (1 AMPw) Fuse (1 AMPw) DC Power Cable: Connect to LCD Module DC Power Cable: Connect to Annunciator Module 20 Amico Pipeline
Appendix C To g A 11 C S 5 up to p 22 ly 0 VA C gl G eS wi tc h N L WIRING DIAGRAM: LCD DISPLAY MODULE - ALARM BUZZER NO COM NC 12V GND NO COM NC 12V GND D C C An on Pow nu ne er nc ct C ia to ab to le : rM od ule NO COM NC 12V GND NO COM NC 12V GND NO COM NC 12V GND NO COM NC 12V GND NOTE: Relays on the annunciator are fail safe for Version 3.1 or newer. SELECT/MUTE CHANGE DOWN CHANGE UP SETUP B4 B3 B2 B1 Relays are not fail safe for Versions 3.0 or older. www.amico.
Appendix D WIRING DIAGRAM: LCD DISPLAY MODULE - LOCAL SENSOR To g A 11 C S 5 up to p 22 ly 0 VA C gl tc h OXYGEN Sensor Module D C C An on Pow nu ne er nc ct C ia to ab to le : rM od ule CHANGE DOWN CHANGE UP SETUP Amico Pipeline Red SELECT/MUTE 22 # 22 Gauge twisted #22 Gauge twisted pair cable cable pairshielded shielded 6” 8” [0.15m 0.2m]supplied 6”-8” [0.1m-0.
Appendix E To g WIRING DIAGRAM: LCD DISPLAY MODULE - REMOTE SENSOR A 11 C S 5 up to p 22 ly 0 VA C gl G eS wi N tc h L # 22 Gauge twisted pair shielded cable 6”-8” [0.1m-0.2m] supplied OXYGEN Fu s e( #22 Gauge twisted pair shielded cable 100’ [30.5 m] max. D C C An on Pow nu ne er nc ct C ia to ab to le : rM od ule Black Red Note: For multiple sensors, a multi-conductor twisted pair cable can be used.
Appendix F WIRING DIAGRAM: LCD DISPLAY MODULE - MASTER MODULE To g A 11 C S 5 up to p 22 ly 0 VA C gl eS wi t#22 Gauge twisted ch L pair shielded cable 6"-8” [0.1 m-0.2 m] supplied OXYGEN Sensor Module G N DISS #22 Gauge twisted pair shielded cable 100’ [30.5 m] max. L N -L G - N ive - G eu ro tra un l d Junction Box (by others) Remote location Alarm D C C An on Pow nu ne er nc ct C at to ab le twisted #22 iGauge or : M od pair shielded ule cable 5,000' [1,500m] max.
Appendix G TECHNICAL SPECIFICATIONS Supply Voltage: 115 - 220 VAC, 50 - 60 Hz Current Draw: 1 Amp. Max. Fuse (1/4 * 1-1/4): Fast Blow 1 Amp. Cable requirement: LCD Alarm to Remote Sensor: Important: Cable: Belden # 8451 or equivalent. #22 gauge shielded, twisted pair. (For multiple sensors a multi-conductor twisted pair, shielded cable should be used). Maximum 100 ft [30.5 m] Distance: Signal: 30 VDC - 1.0 Amps. 60 VDC - 0.3 Amps. 0.5 Amps.
Amico Pipeline
www.amico.
www.amico.com Amico Pipeline | www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.462.6426 Toll Free Fax: 1.866.440.4986 Tel: 905.764.0800 Fax: 905.764.0862 Email: info@amico.com A3-MAN-LCD-ENG 11.12.