User manual

43
2 Select sub menu Set Averaging and press OK. The current
filter constant is shown. Valid values are 0.02, 0.04, 0.08, 0.16,
0.32, 0.64, 1.3, 2.6, 5 and 10 seconds.
3 Change the setting if desired and press
OK.
B.2.12 Setup lamp run time
When the UV lamp is replaced, the Lamp Run Time counter should
be reset.
1 Select sub menu Setup UV and press OK.
2 Select sub menu
Setup Lamp Run Time and press OK.
3 Set the Lamp Run Time counter to zero with the dial. Press OK
to acknowledge.
B.2.13 Setup show UV
Normally UV absorbance is displayed in main menu 1. If not
required, the UV absorbance display can be set to off.
1 Select sub menu Setup UV and press OK.
2 Select sub menu
Setup show UV and press OK. The current
status for showing UV is displayed. If on is shown, the current
UV value is displayed in main menu 1. If off is shown, no UV
value is displayed in main menu 1.
3 Change the setting if desired and press
OK.
B.2.14 Setup language
The language used on the display can be changed.
1 Select main menu
Setup and press OK.
Setup Show UV
(on) on off
Setup Show UV
(on)
Setup
Temp Cond pH UV
Setup Lamp Run Time
Hg (2000h) 000
Setup Lamp Run Time
Hg (2000h)
Setup
Temp Cond pH UV
Set Averaging
(1.3S) 0.64
Set Averaging
(1.3S) 1.3
Monitor UPC-900
User Manual 18-1125-55 Edition AE
Reference information B