User manual
42
B Reference information
3 Change the setting if desired and press OK.
B.2.9 Setup pH temperature compensation
The relationship between pH and the output signal from the pH
electrode is temperature dependent. For accurate measurements
during temperature changes, the pH measurement can be
temperature compensated. In normal applications, when the
temperatures of the buffers and calibration buffers are identical,
temperature compensation is not necessary.
When using temperature compensation, it is important that the
temperature of the pH electrode is the same as that of the
conductivity flow cell since that is where the temperature is
measured.
1 Select sub menu
Setup pH and press OK.
2 Select sub menu
Setup pH Temp Comp and press OK. The
current status for showing pH is displayed. If on is shown,
T
c is
displayed in main menus 1 and 2. If off is shown (default),
T
c is
not displayed in the main menus.
3 Change the setting if desired and press
OK.
B.2.10 Setup show pH
Normally the pH is displayed in the main operating menus (see
section 3.13 Reading pH, temperature and conductivity values). If
not required, the pH display can be set to off.
1 Select sub menu Setup pH and press OK.
2 Select sub menu
Setup Show pH and press OK. The current
status for showing pH is displayed. If on is shown, current pH
is displayed in main menus 1 and 2. If off is shown, no pH is
displayed in the main menus.
3 Change the setting if desired and press
OK.
B.2.11 Setup UV averaging filter constant
1 Select sub menu Setup UV and press OK.
Setup
Temp Cond pH UV
Setup Show pH
(on) on off
Setup Show pH
(on)
Setup
Temp Cond pH UV
Setup pH Temp Comp
(off) on off
Setup pH Temp Comp
(off)
Setup
Temp Cond pH UV
Setup Show Cond
(on) on off
Monitor UPC-900
User Manual 18-1125-55 Edition AE