B.2.7 Setup adjust cell constant
After replacing the flow cell, the cell constant has to be set. (The cell
constant is shown on the packaging).
1 Select sub menu
Setup Cond and press OK.
2 Select sub menu
Setup Adjust Cell Const and press OK.
The current cell constant is shown.
3 A warning message is shown until confirmed by pressing
OK.
4 The current cell constant value is displayed as default. Enter
the new cell constant as read from the packaging and press
OK.
The range is 0.1-300.0 cm
-1
.
When entering the cell constant from UNICORN, select
System
Control:System:Calibrate
and select Cond_Cell.
B.2.8 Setup show conductivity
1 Select sub menu Setup Cond and press OK.
2 Select sub menu
Setup Show Cond and press OK. The current
status for showing conductivity is displayed. If on is shown,
current conductivity is displayed in main menus 1 and 2. If off is
shown, no conductivity is displayed in the main menus.
Setup Show Cond
(on)
Setup
Temp Cond pH UV
Setup Adj Cell Const
(83.56
-1
) 86.6
Warning! This will
change cell calib
Setup Adj Cell Const
(83.56cm
-1
)
Setup
Temp Cond pH UV
Monitor UPC-900