B.2.3 Setup show temperature
The display of the temperature in the conductivity flow cell, shown
in the main operating menu 2, can be enabled or disabled.
1 Select sub menu Setup Temp and press OK.
2 Select sub menu
Setup Show Temp and press OK. The current
status for showing temperature is displayed. If on is shown,
current temperature is displayed in main menu 1 and 2. If off is
shown, no temperature is displayed in the main menus.
3 Change the setting if desired and press
OK.
B.2.4 Setup conductivity temperature compensation
1 Select sub menu Setup Cond and press OK.
2 Select sub menu
Setup Cond Temp Comp and press OK. The
current temperature compensation factor is shown. 0.0% means
that the compensation is off (default setting). The range is
0.0-9.9%.
3 The current compensation factor is displayed as default. Adjust
the compensation factor setting as necessary and press
OK.
B.2.5 Setup conductivity reference temperature
1 Select sub menu Setup Cond and press OK.
2 Select sub menu
Setup Cond Ref Temp and press OK. The
current reference temperature value is shown. 25.0 °C is the
default setting. The range is 0.00-99.9 °C.
3 The current reference temperature value is displayed as default.
Adjust the reference temperature value setting as necessary and
press
OK.
B.2.6 Setup adjust conductivity
Normally it is not necessary to adjust the cell constant as the flow
cell is pre-calibrated on delivery. Adjustment is only necessary when
replacing the conductivity flow cell with a flow cell whose cell
constant is unknown. We recommend that the conductivity flow cell
is recalibrated after cleaning. When adjusting the cell constant from
UNICORN, select System Control:System:Calibrate and then select
CondCalib.
Note: The conductivity temperature compensation must not be used
when adjusting the cell constant. Set the
Setup Cond Temp Comp to
0 (see section B.2.4). The temperature sensor must be calibrated
Setup Cond Ref Temp
(25.0%) 25.0
Setup Cond Ref Temp
(25.0°C)
Setup
Temp Cond pH UV
Setup Cond Temp Comp
(0.0%) 0.0
Setup Cond Temp Comp
(0.0%)
Setup
Temp Cond pH UV
Setup Show Temp
(on) on off
Setup Show Temp
(on)
Setup
Temp Cond pH UV
39
Monitor UPC-900
User Manual 18-1125-55 Edition AE
Reference information B