User manual
38
B Reference information
3 The service telephone number is displayed, press OK.
4 The service contract number is displayed, press OK.
5 The module serial number is displayed, press
OK.
6 The module name and software version are displayed, press
OK.
7 The date of the last service is displayed, press
OK.
8 A test of the module’s buzzer is performed, press
OK.
B.2 Setup menus
B.2.1 Selecting setup sub menus
There are setup sub menus for temperature, conductivity, pH and
UV absorbance.
1 Select menu
Setup and press OK.
2 Select sub menus
Temp, Cond, pH or UV with the dial and press
OK.
B.2.2 Setup adjust temperature
Calibration of the temperature sensor in the conductivity flow cell is
only necessary if the monitor is used in high accuracy measurement
or if the conductivity flow cell is replaced.
1 Place the flow cell together with a precision thermometer inside
a box or empty beaker to ensure that they are not exposed to
draught. Leave them for 15 minutes to let the temperature
stabilise.
2 Read the temperature on the thermometer.
3 Select sub menu
Setup Temp and press OK.
4 Select sub menu
Setup Adjust Temp and press OK. The current
temperature is shown.
5 A warning message is shown until confirmed by pressing
OK.
6 The current adjustment value is displayed as default. Enter the
temperature shown on the thermometer and press
OK.
Setup Adjust Temp
(22.3°C) 22.3
Warning! This will
change temp calib
Setup Adjust Temp
(22.3°C)
Setup
Temp
Cond pH UV
Setup
Temp Cond pH UV
Buzzer Test
Date of Maintenance:
?
UPC-900
V1.00
Serial Number:
01234567 YM 012345
Contract Number:
012345678901
Telephone Service:
012345678901
Monitor UPC-900
User Manual 18-1125-55 Edition AE