User manual
4.8 Cleaning the pH electrode
Note: The pH electrode has a limited lifetime and should be replaced
every six months, or when the response time is slow or the slope is
out of range (<80%).
Use one of the following procedures to clean the electrode to
improve the response:
• Salt deposits: Dissolve the deposit by immersing the electrode
first in 0.1 M HCl, then in 0.1 M NaOH, and again in 0.1 M
HCl. Each immersion is for a 5 minute period. Rinse the
electrode tip in distilled water.
• Oil or grease films: Wash the electrode tip in a liquid detergent
and water. If the film is known to be soluble in a particular
organic solvent, wash with this solvent. Rinse the electrode tip
in distilled water.
• Protein deposits: Dissolve the deposit by immersing the
electrode in a 1% pepsin solution in 0.1 M HCl for five
minutes, followed by thorough rinsing with distilled water.
If these procedures fail to rejuvenate the electrode, the problem is
most likely a clogged liquid junction.
1 Heat a 1 M KNO
3
solution to 60 - 80 °C.
2 Place the electrode tip in the heated KNO
3
solution.
3 Allow the electrode to cool while immersed in the KNO
3
solution before re-testing.
If these steps fail to improve the electrode response, replace the
electrode.
WARNING! HCl and NaOH are injurious to health.
Avoid spillage.
Monitor UPC-900
User Manual 18-1125-55 Edition AE
25
Maintenance 4