4 Maintenance
4 Maintenance
4.1 Periodic maintenance
Interval Action (see procedures below)
Every 3 months Check the monitor
Every 6 months Change the pH electrode
or more often if required Clean the UV flow cell
When required Clean the conductivity cell
Clean the pH electrode
4.2 Cleaning and checking the module
Wipe the monitor housing regularly with a damp cloth. Let the
monitor dry completely before use.
Lamp intensity
1 Select menu
Check and press OK.
2 Select menu
Check Lamp Intensity.
If:
R <300 mV for 254 nm
R <150 mV for 280 nm
R <150 mV for 214 nm,
replace the lamp according to Section 4.10 Changing the UV lamp,
or contact Amersham Pharmacia Biotech for lamp replacement.
Lamp run time
1 Select menu
Check and press OK.
2 Select menu
Check Lamp Run Time.
• The lifetime of a Hg lamp at 254 nm, in room temperature is
typically 7000 hours (in coldroom, typically 2000h).
• The lifetime of a Hg lamp at 280 nm, in room temperature is
typically 3500 hours.
• The lifetime of a Zn lamp is typically 2000 hours in room
temperature.
Check Lamp Run Time
Hg 2300h Zn 340h
Check Lamp Intensity
CAUTION! Only spare parts approved or supplied by Amersham
Biosciences may be used for maintaining and servicing the
module.
WARNING! Always disconnect the power supply before
attempting to replace any item on the module during
maintenance.