User manual
Calibrating with the electrode outside the flow cell
When calibrating the electrode out of the flow cell and changing
from one buffer to another, rinse the electrode tip with distilled
water and dab it carefully with a soft tissue to absorb the remaining
water. Do NOT wipe the electrode as this may charge it and give
unstable readings.
The steps below describe the procedure used with the electrode
removed from the flow cell.
Note: The user interface of the monitor must be unlocked if
connected to a UNICORN control system.
1 Remove the pH electrode from the flow cell and immerse the
electrode in the first standard buffer solution (normally pH 7.0).
2 From main menu 1, turn the dial four steps clockwise to
enter menu
Set Parameters. Press OK.
3 Select
Set Parameters pH . Press OK.
4 Select menu
Calibrate pH. Current calibration values are
displayed (buffer 1 - buffer 2).
Buffer 1=fixed lower calibrated pH value. Range=0.00-14.00
Buffer 2=fixed higher calibrated pH value. Range=0.00-14.00
Note: The pH values for buffer 1 and 2 must differ by at least 1 pH
unit.
5 Press
OK to access the settings menu. The order of calibration,
buffer 1 or buffer 2, is optional. Press OK to start with buffer 1,
or turn the dial to start with buffer 2. In this example, we start
with buffer 1.
6 This text disappears when the reading is stable and the
following text is then shown
:
7 Adjust the pH value in the display using the dial so that it
corresponds to the known pH value of the first buffer solution
and press OK.
8 Turn the dial clockwise to access the buffer 2 calibrating menu.
Rinse the electrode tip with distilled water and then immerse the
electrode in the second buffer solution (e.g. pH 4.0 or 9.0) and
press
OK.
Calibr pH Buffer 2
Calibr pH Buffer 1
Calibr pH Buffer 1
Calibr pH Buffer 1
Calibrate pH
(7.00 - 12.00)