User manual

2 Installation
2.1 Unpacking
Unpack the module and check the items against the supplied packing
list. Inspect the items for obvious damage that may have occurred
during transportation.
Note: It is important that the filters, flow cells and lamps are not
handled during unpacking. For protection of these items, they should
remain in their packing materials until required for use.
We recommend that all packing materials be retained if onward
transport of the module is expected.
Note: pH measurement is optional in ÄKTA
FPLC.
2.2 General precautions
The module should be located in a place of low temperature
variations, away from heat sources, draughts and direct sunlight.
The module may be operated at normal ambient temperatures in
the range +4 to 40 °C.
The module should be installed on a stable laboratory bench or
in ÄKTA
FPLC
system rack. To ensure correct ventilation, a free
space of 0.1 m is required behind and in front of the module. Do not
place soft material under the module to ensure that the ventilation
inlet below the front is not blocked.
UPC-900
CAUTION! Read the following information carefully to ensure
that the module is installed correctly.
Monitor UPC-900
User Manual 18-1125-55 Edition AE
3
Installation 2