Manual

2
Introduction
Model 286 Features
indicates liquid level and fill/loss rate. The current liquid level is updated
and visible in many of the display modes used in configuring various
settings of the instrument.
The 4 x 4 keypad is provided for direct entry of values for setpoints, sensor
length, and other functions. The keypad also provides a convenient menu-
type interface for configuring various instrument settings that require
selection from a list of options.
1.1.4 HI/LO Alarms
Two user-defined setpoints, HI and LO, are provided for all four channels
with front panel LED indications for Channels 1 and 2. The HI and LO
setpoints may also operate two pairs of low-current, rear panel relay
contacts and a built-in audible alarm. The HI/LO relay contact pairs may
be assigned to Channels 1 and 2, or to Channels 3 and 4.
1.1.5 Controller Modes
The Model 286 provides three controller modes: normal, auto-changeover,
and pre-cool. The preferred mode can be selected via the front panel
keypad or remote interface. The level control sequence varies based on the
selected mode. See the discussion beginning on page 7 for more
information regarding the function of the controller modes.
1.1.6 Remote Operation
The Model 286 is capable of providing up to two analog 010 VDC signals,
which corresponds to 0100% liquid level, accessible from the rear panel of
the instrument for use with a voltage recorder. This is a factory-installed
option and should be specified at the time the instrument is purchased.
Up to two self-powered (i.e. does not require an external power supply) 4-
20 mA current loop outputs can also be ordered as an option.
Additional digital interface options, including RS-232 or RS-422, are
available as options. The serial remote interface provides a comprehensive
command set for remote monitoring and configuration of all functions of
the instrument.