37
Remote Interface Reference
Error Codes
4.3 Error Codes
The Model 135/136 returns specific error codes for invalid commands and/
or arguments. If an error condition is returned, the command is not
processed and the configuration of the instrument is
not
modified. The
table below provides a list of error codes, their meaning, and any
associated limits. The error codes are common to both the serial and IEEE-
488 communication options.
Error Code Meaning Valid Range
1
LO setpoint out of range
0
LO
LENGTH
2
B setpoint out of range
0
B
<
A
3
A setpoint out of range
B < A
LENGTH
4
HI setpoint out of range 0
HI
LENGTH
5
Attempted to set or query for
LENGTH in PERCENT units mode
6
Invalid argument,
value
out of
maximum calibration range
1 cm
value
203.2 cm
7
INTERVAL setting out of range
0
INTERVAL
600 min
8
Unrecognized command
9
Invalid argument, value was negative
or non-numeric