Manual

34
Rev. 3
Remote Interface Reference
IEEE-488 Command Set Reference
the sampling interval to 0.1 minutes (or 0.1 hours if so configured, see
page 53). The SAVE command saves the HI, LO, A, B, INTERVAL,
LENGTH, and current remote units settings to permanent memory. These
settings are then recalled each time the power is turned off and then
reapplied to the instrument. If the configuration is changed from the front
panel, the settings are automatically saved to permanent memory.
4.2.4.4 Commands for querying the configuration
The LENGTH, HI, LO, A, B, and INTERVAL commands return the
current configuration of the instrument. Each return value is terminated
with
<LF>
and EOI.
Command: LENGTH Function: Returns the sensor
length in the
current units
Returns: <value>
Command: HI Function: Returns the HI
setpoint limit in the
current units
Returns: <value>
Command: LO Function: Returns the LO
setpoint limit in the
current units
Returns: <value>
Command:
A
Function: Returns the A
setpoint limit in the
current units
Returns: <value>
Command:
B
Function: Returns the B
setpoint limit in the
current units
Returns: <value>
Command: INTERVAL Function: Returns the
sampling interval in
the current units
Returns: <value>