Remote Interface Reference
IEEE-488 Communication DIP Switch Settings
4.2.3 IEEE-488 Communication DIP Switch Settings
The 5 DIP switches located on the rear panel of the Model 135/136 are
used to control the IEEE-488 interface. Switches 1 through 5 control the
IEEE-488 bus address of the unit.
4.2.3.1 IEEE-488 primary bus address
The Model 135/136 primary bus address is controlled by switches 1
through 5 of the communication DIP switch on the rear panel. Valid
primary addresses are between 0 and 30. The Model 135/136 does not use
secondary addressing. Note that many IEEE-488 controller cards in
external computers will use address 0. The bus address for each Model
135/136
should be unique
with respect to other Model 135/136 units or any
other devices on the bus. The switch settings for the various addresses are
(on = 1 or the up position):
DIP switch Primary bus
address
12345
off off off off off 0
off off off off on 1
off off off on off 2
off off off on on 3
off off on off off 4
off off on off on 5
off off on on off 6
off off on on on 7
off on off off off 8
off on off off on 9
off on off on off 10
off on off on on 11
offononoffoff 12
off on on off on 13
offonononoff 14
offonononon 15
on off off off off 16
on off off off on 17
1
0
12345