Installation Guide

98 8200-1953-09 A0
of RF radiation equipment.
Electrical fast tran-
sients / Bursts IEC
61000-4-4
100 kHz repetition fre-
quency ± 2 kV for AC
and DC power input ±
1 kV for input / output
lines
Complies
AC and DC power quality should be
that of a typical commercial or home
healthcare, environment.
Surge IEC 61000-4-5
± 0,5 kV, ± 1 kV Line-
to-line (AC and DC) ±
0,5 kV, ± 1 kV, ± 2 kV
Line-to-ground (AC
and DC) ± 2 kV Line-
to-ground Input/output
lines
Complies
AC and DC power quality should be
that of a typical commercial or home
healthcare, environment.
Conducted dis-
turbances induced by
RF fields IEC 61000-4-
6
10 V a) 0,15 MHz 80
MH 6 V in ISM and
amateur radio bands
between 0,15 MHz
and 80 MHz 80 % AM
at 1 kHz
Complies
Portable and mobile RF com-
munications equipment should follow
the guidance in the table titled Recom-
mended separation distances between
portable and mobile RF com-
munication equipment and the medical
equipment.
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey should be less
than the compliance level in each fre-
quency range.
Interference may occur in the vicinity
of RF radiation equipment.
Rated power frequency
magnetic fields IEC
61000-4-8
30 A/m 50 Hz and 60
Hz
Complies
Power frequency magnetic fields
should be that of a typical commercial
or home healthcare, environment.
Voltage Dips IEC
61000-4-11
0 % UT; 0,5 cycle At
0°, 45°, 90°, 135°,
180°, 22, 270° and
315° 0 % UT; 1 cycle
and 40% UT; 10/12 at
50/60Hz b) and 70 %
UT; 25/30 cycles
Single phase: at 0°
Complies
AC and DC power quality should be
that of a typical commercial or home
healthcare, environment. If the med-
ical equipment requires continued oper-
ation during power interruptions, then it
is recommended that the equipment be
powered from an uninterruptable power
supply.
Voltage Interruptions
IEC 61000-4-11
0 % UT; 250/300 cycle Complies
AC and DC power quality should be
that of a typical commercial or home
healthcare, environment.
a) Voltage level increased to reflect requirement in EN 61000-6-2.