Installation Guide

96 8200-1953-09 A0
EMC Information
ME system is not specified for use only in a shielded location.
(Need statement(s) of the environments for which the ME system is suitable.)
Essential performance of the medical system exists when video monitoring is carried out in
the designated area to detect the body temperature within the monitoring range of the
camera in real time with high precision. The monitoring images and body temperature can
be viewed in real time through the web client on a PC. Essential performance may be lost or
degraded due to electromagnetic (EM) disturbance. The system operator may experience
one or more of the following conditions due to degraded or lost essential performance.
Temporary loss or drop out of video
Video flicker but
System reboot
Variation in set temperature (< 0.1°C)
Loss of alarm functions
The system resumes normal operation after the EM disturbance is removed.
Avoid stacking or placing equipment adjacent with other equipment to prevent improper
operation of the products.
The use of accessories, transducers, and cables, other than those specified or provided by
the manufacturer could result in increased electromagnetic emissions or decreased
electromagnetic immunity and result in improper operation.
Portable RF communications equipment, including peripherals such as antenna cables and
external antennas should be no closer than 12 inches (30cm) to and part of the medical
system, including cables specified by the manufacturer.