Installation Guide

Tyco Illustra Pro 5MP Thermal Elevated Skin Temperature Detection Camera Installation and Configuration
8200-1953-09 A0 44
Note:A red box is now visible in the thermal image window. See example in Figure
28. The blackbody should not be visible within this area for accuracy measurement.
If the blackbody is outside of the box then move it within the box.
Note:It is important that the blackbody is in a position where it will not be obscured
during detection.
Figure 29 'Red Box' visible in the thermal image window
8 Select the Blackbody Temperature text box and enter a value.
It is recommended that this is set to 35°C
Note:Turn on the blackbody and then press Up or “Down” to increase or decrease
the temperature value. It can take about 20~30 minutes for the blackbody to be
steady.
9 Select the Blackbody Mode drop-down menu and select Auto or Manual.
Note:It is recommended to select Auto” mode and use the default settings.
Note:If Manual” is selected, then please calibrate the temperature as follows.
Move the cursor to the blackbody image in the temperature calibration interface.
The temperature is then visible on screen. If this value is not 35°C, please select
the Temperature Correction text box and enter the value.
10 Select the Temperature Correction text box and input the value offset from the overlay
blackbody temperature.
Example: If The Blackbody is set to 35.0 and the camera is reading 34.9 then the offset
would be 0.1.
Note:Configure and save the calibration settings from this procedure before
establishing this offset.
11 Select the Calibarate button and then select Save to save the settings.