Tyco Illustra Pro 5MP Thermal Elevated Skin Temperature Detection Camera Installation and Configuration
8200-1953-09 A0 17
• Elevated skin temperature in the context of use should be confirmed with a secondary
evaluation method such as a clinical grade contact thermometer.
• The system operator must ensure the face is unobstructed by hair, eyeglasses, and
other objects because their presence will interfere with the ability of a measurement
thermograph to detect a febrile condition.
• The relative humidity in the measurement area should be maintained below 50 % and
the temperature below 24 °C for best performance. A measurement area with elevated
humidity and ambient temperature may lead to inaccurate temperature measurements
due changes in skin temperature caused by sweating.
• Public health officials, through their experience with the device in the particular
environment of use, should determine the significance of any fever or elevated
temperature based on the skin tele thermographic temperature measurement.
• This device should be used to measure only one subject's temperature at time
in accordance with IEC 80601-2-59:2017 and FDA (or other applicable
regulatory) guidelines for accuracy.
• The visible thermal patterns of this device are only intended for locating the
points from which to extract the thermal measurement.
• The product calibration interval is 14 days.
Factors to consider for the environment and installation that can
impact measurement
• Thermal camera is temperature-sensitive. When it is used for temperature measuring,
in order to ensure the accuracy, the recommended ambient temperature for operation is
0~35°C and the infrared light should be turned off during the temperature measurement.
• Thermal camera should always be used together with blackbody both for increased
accuracy and to comply with IEC standards for measurement thermographs for human
febrile temperature measurement.
• Thermal camera should be used in a stable indoor environment without wind. Please
make sure the monitoring field is far away from any objects that could produce airflow,
high temperature, reflection.
• Don’t install the thermal camera opposite to a door, air conditioner or any place in or
near infrared sources such as sunlight.
• For monitoring, the best temperature measurement distance is 2~4m, recommended
distance should be 2~6m to achieve specified accuracy.
• Set up a one-way temperature measuring zone and make sure the camera can clearly
see the human face.
• After turning on, the thermal camera needs to wait 20~30 minutes to be steady.
• It is recommended to use DC 12V power supply to reduce power consumption.
• The blackbody used with the camera shall be installed 2 or 3 meters away from the
camera and opposite to the camera. It cannot be blocked. During operation, do not
move the camera and blackbody and make sure the blackbody is in a fixed and proper
place in the camera image for calibration and thermal drift compensation.
• The thermal camera and the blackbody device should be at a height of 2-3m and
parallel to the faces being screened for temperature.