Calibration / accuracy
This device is specified to achieve a +/- 0.3°C accuracy when configured properly
following the installation guide and calibrated against a blackbody device. Calibration
must be done following device calibration intervals in line with the user manual
guidance in order to maintain the specified accuracy.
The blackbody device is used to manage thermal drift compensation and is crucial in
obtaining accurate temperature assessments in line with the product accuracy
specifications.
The solution uses the forehead region of the face using 4 sampling spots of 3x3 pixels
as the workable target plane. The total of 36 pixels is then verified to be within 0.2°C
averaged across all the pixels to ensure consistency with the highest observed
temperature being reported with the camera conducting automated calibration with the
blackbody device to manage thermal drift compensation.
Guidelines
System architects, installers, users, and operators should refer to guidelines in ISO/TR
13154: Medical electrical equipment Deployment, implementation and operational
guidelines for identifying febrile humans using a measurement thermograph.
Temperature measurement with this device should not be solely or primarily relied upon
to diagnose or exclude a diagnosis of any illness, disease or other medical condition.
The device is tested and labelled consistent with the standard IEC 80601-2-59:2017:
Medical electrical equipment Part 2-59: Particular requirements for the basic safety
and essential performance of measurement thermographs for human febrile
temperature measurement.