100 8200-1953-09 A0
Recommended separation distances between portable and mobile RF communication
equipment and the medical equipment
The medical equipment is intended to be installed in an environment in which radiated RF disturbances are
controlled. To minimize electromagnetic interferences, the user should maintain distances from portable
and mobile RF communications equipment (e.g. transmitters) and the medical equipment and its cables as
shown below. The user should also ensure that maximum power, as shown below, from nearby RF
devices is not exceed.
Test Fre-
quency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385 380 – 390 TETRA 400
Pulse mod-
ulation 18 Hz
1,8 0,3 27
450 430 – 470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation 1
kHz sine
2 0,3 28
710
704 – 787
LTE Band 13,
17
Pulse mod-
ulation 217 Hz
0,2 0,3 9745
780
810
800 - 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse mod-
ulation 18 Hz
2 0,3 28870
930
1720
1700 -
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse mod-
ulation 217 Hz
2 0,3 281845
1970
2450
2400 –
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse mod-
ulation 217 Hz
2 0,3 28
5240
5100 -
5800
WLAN
802.11 a/n
Pulse mod-
ulation 217 Hz
0,2 0,3 95500
5785