Desk Charging Base Instruction Manual Questions? Call ADC toll free: 1-800-232-2670
1. A SPECIAL THANK YOU Congratulations on your purchase of an ADC® Diagnostix™ Desk Charging Base. ADC® professional diagnostic products are the instruments of choice where accuracy and dependability are critical. This feature rich instrument was designed to simplify physical exams and noninvasive diagnostics with proper use and care it will provide many years of dependable service. Read this booklet thoroughly before attempting to use your new ADC® Diagnostix™ Desk Charging Base. 2.
Warning: The device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g., mobile telephones, microwave ovens). These can lead to temporary impairment of the ADC® Diagnostix™ Desk Charging Base. Warning: Do not use batteries, electrical cords other than those included with this product or replacement parts supplied by the manufacturer.
c) Charging Time • The end of the charging time is detected by evaluation of the negative voltage differential caused by heating of the rechargeable battery cells when fully charged. The yellow charging LED will switch off and the green ”rechargeable battery full” LED will switch on. • The charging time is limited to max. 4.75 h by the internal timer. • When the rechargeable battery is fully charged (green LED on), the unit will switch to maintenance charge.
6. WALL MOUNTING The ADC® Diagnostix™ Desk Charging Base may be mounted on the wall. Hold wall rail to the place on the wall intended for mounting. Mark holes with a pen and drill holes of 5 mm (3/16”). Insert the dowels supplied. Position rail holes on the dowels and fix to the wall with the screws supplied. Then slide unit on bracket to allow the wall rail to be introduced into the two guideslots of the ADC® Diagnostix™ Desk Charging Base plate.
Sterilization According to the Test Centre for Medical Devices in Tübingen, Germany, sterilization is only prescribed in the case of operative procedures. There is no necessity for sterilisation of this charging base. 9. TECHNICAL SPECIFICATIONS Model: ADC® Rechargeable Battery Charger Base Display System: LEDs Connection Main Connection (See Note on Bottom Label) Output: 2.4 V (d.c.
12. WARRANTY This ADC® Diagnostix™ Desk Charging Base is warranted for 2 years from date of purchase. This warranty includes the instrument. The warranty does not apply to damage caused by improper handling, accidents, improper use, or alterations made to the instrument by third parties. The warranty is only valid after the product is registered online at www.adctoday.com. 13. QUALITY STANDARDS Electromagnetic compatibility: Device fulfills the stipulations of the International standard IEC60601-1-2 14.