Mechanical thermostat sealed in UL listed explosion proof box
Adjustable operating temperature range:
Refrigerator: 1°C 10°C
Freezer: -15°C -25°C
Keyed door lock
Magnetic door gasket for positive seal
Heavy duty door handle
Leveling legs
White textured cabinet and door
High density urethane foam cabinet and door insulation
CFC-free refrigerant, foam insulation and packaging
Defrost:
Refrigerator: Cycle defrost
Freezer: Manual defrost
Not equipped with line cord and plug; must be hardwired in conduit to 120
VAC power source
UL/C-UL listed
One year parts and labor warranty
Meet NFPA and OSHA guidelines for hazardous locations, Class 1, Division
II, Group C & D.
These units are not intended for use in general clinical or research
laboratories.
Standard Hazardous Location
(Explosion Proof)
Refrigerators & Freezers
Standard Explosion Proof
Hazardous Location
Refrigerators & Freezers
Standard Hazardous Material Storage (Explosion Proof)
Refrigerators & Freezers meet NFPA and OSHA guidelines
for Hazardous Locations, Class 1, Division II, Group C & D.
These units are engineered for use in environments where
volatile / explosive conditions could potentially exist.
They are not intended for common lab / room environments
that are not designed to be explosion proof (See: Flammable
Material units).
Units are designed with no internal electrical components
and all external components are sealed per NFPA and OSHA
guidelines. Units must be hard wired in conduit to 120VAC
power source.
For manual defrost freezers, periodic manual defrosting is
required.
Power cords with plugs are NOT ACCEPTABLE.
Compact Laboratory
Refrigerators
Refrigerators
Model Cu. Ft. Defrost Doors Int Door Shelves W” D” H” RL H.P. Amps
ABT-ERS-14 14 Cycle 1 Solid 0 1 Adj/2 Fixed 29 34 ½ 66 ½ R134a 5
ABT-ERS-17 17 Cycle 1 Solid 0 1 Adj/2 Fixed 34 32 ½ 72 R134a 5
ABT-ERS-20 20 Cycle 1 Solid 0 1 Adj/3 Fixed 34 34 ½ 77 ¾ R134a 5
ABT-ERS-30 30 Cycle 1 Solid 0 5 Adj 35 ¾ 35 77 ½ R404a
6
ABT-ERCS-30 30 Cycle/Manual 1 Solid 1 Freezer 3 Adj-Ref 35 ¾ 35 77 ½ R404a
6
Freezers
Model Cu. Ft. Defrost Doors Int Door Shelves W” D” H” RL H.P. Amps
ABT-EFS-14 14 Manual 1 Solid 0 3 Fixed 29 34 ½ 66 ½ R134a 5
ABT-EFS-17 17 Manual 1 Solid 0 3 Fixed 34 32 ½ 72 R134a 5
ABT-EFS-20 20 Manual 1 Solid 0 4 Fixed 34 34 ½ 77 ¾ R134a 5
ABT-EFS-30 30 Manual 1 Solid 0 4 Fixed 35 ¾ 35 77 ½ R404a
6
Mechanical thermostat
sealed in UL listed
explosion proof box
Meets NFPA and OSHA
guidelines for hazardous
locations, Class 1,
Division II, Group C & D
ABT-ERS-20
ABT-EFS-30