Brochure
ABT-FRP-04
Both units feature digital
displays, alarms and
remote alarm contacts
ABT-HC-FRFC-10
52
• Microprocessor Temperature Controller
• Adjustable operating temperature range:
Refrigerator: 1°C – 10°C
Freezer: -15°C – -25°C
• Digital Temperature Display
• Remote alarm contacts
• Adjustable temperature display in Celsius or Fahrenheit
• Keyed door lock
• Audible and visual high/low temperature alarms
• Magnetic door gasket for positive seal
• Integrated door handle
Premier Flammable Storage
Refrigerators & Freezers with Natural
Refrigerants
The Premier Flammable Storage Refrigerators and Freezers with
Natural Refrigerants are purpose built for scientific and medical
applications. These units meet NFPA and OSHA guidelines 45 and
70 for the refrigerated storage of flammable (volatile) materials in
general laboratory work areas.
The all-new temperature display and alarm module offers product
security with a monitoring and alarm system.
Real-time sample simulated temperatures are constantly
displayed and supported by audible/visual high and low
temperature alarms. Remote alarm dry contacts and an access
port are provided to simplify connections to existing facility
monitoring systems.
For manual defrost freezers, periodic manual defrosting is
required.
There are no electrical components located in the interior of
the unit and the compressor’s electrical components have been
sealed in a vapor-proof enclosure for additional safety.
These units are NOT designed for use in environments where
volatile / explosive conditions could potentially exist.
Premier Flammable Storage
Refrigerators & Freezers with
Natural Refrigerants
53
• Defrost:
Refrigerator: Cycle defrost
Freezer: Manual defrost
• UL/C-UL listed
• Units meet NFPA guidelines 45 and 70
• Leveling legs
• One probe access port (⅜”)
• White textured cabinet and door
• High density urethane foam cabinet and door insulation
• CFC-free refrigerant (R134a), foam insulation and packaging
• Two year parts and labor warranty
ABT-HC-FFP-20
Digital temperature
display and alarm
module.
Refrigerators
Model Cu. Ft. Defrost Doors Shelves W” D” H” RL H.P. Amps
ABT-FRP-04 5 Cycle 1 Solid 3 Adj 23 ⅝ 24 ¼ 32 / R600a 1/9 1.2
Refrigerators/Freezer Combos
Model Cu. Ft. Defrost Doors Shelves W” D” H” RL
ABT-HC-FRFC10 5/4 Cycle/Manual 2 Solid 3 Adj Ref/2 Adj Fr 23 ⅝ 24 ¼ 65 ⅜ R600a
Freezers
Model Cu. Ft. Defrost Doors Shelves W” D” H” RL H.P. Amps
ABT-FFP-04 4 Manual 1 Solid 2 Fixed 23 ⅝ 24 ¼ 32 / R600a 1/8 1.4
ABT-HC-FFP-14 14 Manual 1 Solid 4 Fixed 23 ⅝ 24 ⅜ 73 R290a 1/2 1.96
ABT-HC-FFP-20 20 Manual 1 Solid 6 Fixed 30 ¾ 28 ½ 74 ⅝ R290a 1/2 1.96
The Premier Natural Refrigerant Manual Defrost Laboratory
Freezer takes a large step forward in performance from past solid
door general purpose units; offering a Digital temperature display
and alarm module coupled with a microprocessor temperature
controller.
-20
Freestanding models require 4” of clearance around back and sides of the unit for proper ventilation. Improper
installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure.
Freestanding models require 4” of clearance around back and sides of the unit for proper ventilation. Improper
installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure.
Units meet National
Fire Protection
Association guidelines
45 and 70
Environmentally friendly, natural hydrocarbon (HC) refrigerants which
vastly reduce global-warming potential while meeting new EPA/SNAP