Models/Modelos/Modéles BB15-CO, BB15-CO2, BB30-CO, BB30-CO3, BB50-CO, BB75-CO, BB100-CO, BB100-CO6, BB100-CO8, BB150-CO Manual No. BREBX013 (Rev 6 July 2008) Operating Manual Manual de instrucciones Manuel d’utilisation AIR SYSTEMS INTERNATIONAL, INC. 829 Juniper Crescent, Chesapeake, Va., 23320 Telephone (757) 424-3967 Toll Free 1-800-866-8100 Fax No. (757) 424-5348 http://www.airsystems.com Air Systems International, Inc e-mail: sales@airsystems.com Registered to ISO 9001 Certificate No.
-2- SPECIFICATIONS BB15 SERIES Size BB30 SERIES BB50 SERIES BB75 SERIES BB100 SERIES BB150 SERIES 19" L x 13" H x 7"D 19"L x 13"H x 7"D 23.5"L x 16.75"L x 8.5"D 23.5"L x 16.75" H x 8.5" D 26.25"L x 20.75" H x 9" D 26.25"L x 20.75" H x 9" D Weight Inlet Size 12.3 lbs/5.6 kg 1/4" Industrial Interchange 19.7 lbs/8.0 kg 1/2" Industrial Interchange 2 Standard 3 Optional No. of Outlets 1 Maximum Air 15scfm @ 110psi 30scfm @ 110psi Flow (cfm/bar) 425 lpm @ 7.5 bar 850 lpm @ 7.
-3- BREATHER BOXTM PARTS IDENTIFICATION MODEL BB15-CO MODELS BB30-100 MODEL BB150-CO
-4- BREATHING AIR QUALITY POSITION STATEMENT The responsibility for the quality of breathing air rests with the user. Compliance with federal, state, or local regulations are the responsibility of the user and this recommendation does not supersede any existing rules, regulations, or laws which may apply. Breathing air filtration products meet or exceed CGA Grade-D specifications for air quality as adopted by Federal OSHA. Compressor air quality standards meet or exceed OSHA 1910.134 requirements.
-5- MONITOR OVERVIEW The monitor will analyze the air sample and display the CO concentration in parts-per-million (ppm). The system’s green “NORMAL” operation light will illuminate and the red “HIGH CO” light will flicker faintly approximately every second when the CO level is below 10ppm (5ppm Canadian).
-6- ITEM # 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 DESCRIPTION PART # LCD DISPLAY MONC703 SPAN POTENTIOMETER MONC702A ALARM SET POINT POTENTIOMETER MONC702A ZERO POTENTIOMETER MONC702 AIR SAMPLE INLET SOCKET MONC001 AIR SAMPLE PLUG MONC002 AIR EXHAUST PORT MONC003 ON/OFF/TEST SWITCH MONC007 RECESSED PLUG WITH FUSE HOLDER MONC020 1 AMP FAST ACTING FUSE, 5 X 20MM ELF001 15 PIN SOCKET MONC520 FACEPLATE/ENDPLATE SCREW MONC023 MAIN CIRCUIT BOARD ASSEMBLY CO-91P
-7- BREATHER BOX TM SETUP AND OPERATION Note: Always operate the Breather BoxTM in the upright position. Failure to comply may result in one or all of the following: • Auto drains will not function properly. This may result in the contamination of the CO monitor and cause water to be passed through respirator hose and into the worker’s mask. • Auto drains may become clogged, clean or replace auto drains (See Maintenance Instructions.
-8- STEP 6) Connect the air source to the inlet fitting. MODEL # BB15-CO BB30-CO BB50-CO BB75-CO BB100-CO BB150-CO MIN. HOSE ID 3/8" 1/2" 1/2" 1/2" 1/2" 3/4" 1/4" 1/2" 1/2" 1/2" 1/2" INLET FITTING INDUSTRIAL INTERCHANGE INDUSTRIAL INTERCHANGE INDUSTRIAL INTERCHANGE INDUSTRIAL INTERCHANGE INDUSTRIAL INTERCHANGE 1" CHICAGO FITTING STEP 7) Hold the “ON/OFF/TEST” switch in the “TEST” position. All local and remote audible/visual indicators will activate.
-9- SHUTDOWN 1) Make sure all personnel have egressed from the work area. 2) Shut off air source to the box. 3) Remove air pressure from the box by pulling the relief valve ring out. 4) Turn monitor “OFF” at the “ON/OFF/TEST” switch. Do not remove 9-volt batteries. These are used to maintain a bias voltage to the sensors; this keeps the sensor ready for immediate future use. 5) Disconnect airline hoses. 6) Install dust caps if applicable.
-10- SENSOR REPLACEMENT Replacement sensors are shipped with a metal spring installed between the electrodes. Do not remove the clip until the sensor is to be installed into the monitor. STEP 1) Disconnect all external connections. Remove CO monitor from the unit. STEP 2) STEP 3) Remove the four screws from the monitor’s Remove end plate to gain access left end plate. to the sensor cup from outside the Note: Alarm location may vary. housing. STEP 4) Remove sensor from sensor cup and remove leads .
-11- CALIBRATION PROCEDURE Do not use inert gases to zero the monitor. This will cause premature failure of the sensor. CO Zero Adjustment To zero the instrument, follow the steps below. Zero calibration gas should be used to properly “zero” the instrument and assure that a valid calibration is achieved. If zero adjustment cannot be made as indicated, sensor replacement may be necessary. After each monitor adjustment outlined in the following steps, allow time for the changes to stabilize. 1.
-12- CO SPAN ADJUSTMENT Use only 10 - 20 ppm CO gas for calibration. Using a higher concentration may decrease accuracy at lower scale readings. Note: 10ppm gas must be used to satisfy Canadian calibration requirements. 1. Install regulator to the CO calibration gas cylinder. 5. Adjust the "span" pot adjustment screw (clockwise to increase, counterclockwise to decrease) until the digital display reads the same concentration (ppm) as printed on the calibration gas cylinder. 2.
-13- MODEL BB15-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 7 DESCRIPTION P/N “A” FILTER ELEMENT AND O-RING BB15-AW “C” FILTER ELEMENT AND O-RING BB15-CW “D” FILTER ELEMENT AND O-RING BB15-DW AUTO DRAIN ASSEMBLY 15ADW FILTER BOWL W/AUTO DRAIN 15-PBAW MANUAL DRAIN 15MDW FILTER BOWL W/MANUAL DRAIN 15-PBMW * INCLUDED WITH FILTER ELEMENTS NOTE: FOR UNITS PURCHASED PRIOR TO 12/04, PLEASE CONTACT CUSTOMER SERVICE FOR CORRECT ORDERING INFORMATION ON FILTER REPLACEMENT ELEMENTS.
-14- MODEL BB30-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 7 8 9 DESCRIPTION FILTER CHANGE INDICATOR FILTER BOWL O-RING “A” FILTER ELEMENT “C” FILTER ELEMENT “D” FILTER ELEMENT AUTO DRAIN ASSEMBLY FILTER BOWL W/GUARD MANUAL DRAIN FILTER BOWL W/GUARD & MANUAL P/N WL261 WL266 BB30-A BB30-C BB30-D WL024 WL264 WL262 WL267
-15- MODEL BB50-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 7 8 9 DESCRIPTION FILTER CHANGE INDICATOR FILTER BOWL O-RING “A” FILTER ELEMENT “C” FILTER ELEMENT “D” FILTER ELEMENT AUTO DRAIN ASSEMBLY PLASTIC FILTER BOWL METAL BOWL GUARD MANUAL DRAIN P/N WL056 WL091 BB50-A BB50-C BB50-D WL024 WL049 WL094 WL153
-16- MODEL BB75-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 7 8 DESCRIPTION FILTER CHANGE INDICATOR “A” FILTER ELEMENT AND O-RING “C” FILTER ELEMENT AND O-RING “D” FILTER ELEMENT AND O-RING AUTO DRAIN ASSEMBLY FILTER BOWL W/AUTO DRAIN MANUAL DRAIN FILTER BOWL W/MANUAL DRAIN P/N WL261 BB75-A BB75-C BB75-D WL024 WL187 WL262 WL188 * Must Order the 4-outlet or 6-outlet manifold. Unit comes standard with 1/2" FPT inlet/outlet unless otherwise specified.
-17- MODEL BB100-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 7 8 9 DESCRIPTION FILTER CHANGE INDICATOR FILTER BOWL O-RING “A” FILTER ELEMENT “C” FILTER ELEMENT “D” FILTER ELEMENT AUTO DRAIN ASSEMBLY PLASTIC FILTER BOWL METAL BOWL GUARD MANUAL DRAIN P/N WL056 WL113 BB100-A BB100-C BB100-D WL024 WL055 WL092 WL153
-18- MODEL BB150-CO REPLACEMENT FILTER BREAKDOWN ITEM # 1 2 3 4 5 6 8 9 10 DESCRIPTION FILTER CHANGE INDICATOR FILTER BOWL O-RING “A” FILTER ELEMENT “C” FILTER ELEMENT “D” FILTER ELEMENT AUTO DRAIN ASSEMBLY PLASTIC FILTER BOWL METAL BOWL GUARD MANUAL DRAIN P/N WL056 WL113 BB100-A BB100-C BB100-D WL024 WL055 WL092 WL153
-19- Warranty Disclaimer Air Systems’ manufactured equipment is warranted to the original user against defects in workmanship or materials under normal use for one year after date of purchase. Any part which is determined by Air Systems to be defective in material or workmanship will be, as the exclusive remedy, repaired or replaced at Air Systems’ option. This warranty does not apply to electrical systems or electronic components.