4 CALIBRATION
26
4.4 Manual Calibration
After preparing for the calibration (see “ 4.2 Preparation for Calibration ” (page 14)), perform
the zero calibration and the span calibration in this order.
4.4.1 Operational flow
The operational flow for manual calibration is described below:
Note
Control the zero gas and span gas feed pressures at atmospheric pressure plus (0 kPa to 0.5 kPa).
When using the calibration gas line: When using the measured gas line Reference page
Gas line connection check
↓|
Span value setting page 14
↓↓
Measurement line switching
[MEAS.] -> [ZERO]
Measurement line check
[MEAS.]
page 27
|↓
Zero gas introduction
from the calibration gas line
Zero gas introduction
from the sample inlet
↓|
Waiting for readout stabilization
↓↓
Zero calibration
↓|
Measurement line switching
[ZERO] -> [SPAN]
|
page 28
↓↓
Span gas introduction
from the calibration gas line
Span gas introduction
from the sample inlet
↓↓
Waiting for readout stabilization
↓↓
Span calibration
↓↓
Line switching
[SPAN] -> [MEAS.]
|
↓
Sample gas introduction
from the sample inlet