Product specifications

4 CALIBRATION
22
4.3.3 Setting the AIC sequence
To set the AIC sequence, go to the AIC SEQUENCE screen.
1. Press the [MENU] key on the MEAS. screen.
2. Press the [W] or [X] key to display the MENU/SETTING screen.
Fig. 27 MENU/SETTING screen
3. Press the [AIC SEQUENCE] button. The AIC SEQUENCE screen will be displayed.
Fig. 28 AIC SEQUENCE screen
Item Settable range Description
WAIT 0 min to 999 min
Set the waiting time for stabilization after changing gas.
The recommended setting is 20 minutes or longer.
HOLD 0 min to 999 min
Set the calibration validation time (to check the readouts on the recorder after finishing the
calibration).
CAL. YES/NO
Specify whether or not to perform calibration.
YES: Calibration is performed. NO: Calibration is not performed.