Product specifications

4 CALIBRATION
17
Note
Only when AIC MODE is set to INTERNAL, the items of START TIME, LIMIT (START-END), and
INTERVAL are displayed. These items are not displayed when AIC MODE is set to NONE or
EXTERNAL.
4. Press the displayed item to be set. The corresponding setting screen will be
displayed.
For the detailed explanation of each screen, see page 17 to page 21.
5. On the setting screen, change the settings and then press the [SET] key.
The changed settings will be saved, and the AIC screen will be displayed again.
Tip
To cancel the changes, press the [CANCEL] key. The changes will be undone, and the AIC screen will
be displayed again.
6. Press the [CLOSE] key on the AIC screen to return to the MENU screen.
Q AIC MODE
Specify the method of starting the AIC.
Pressing the displayed AIC MODE setting will display the AIC MODE screen.
Fig. 21 AIC MODE screen
Press the button of the item to be set.
Note
z Manual AIC start and the start via the RS-232C port are valid regardless of this setting.
z If an AIC start signal is inputted externally while an AIC sequence is in progress, this signal will
be disregarded and the ongoing AIC sequence will be continued.
Item Description
INTERNAL Selects the mode of using the internal clock to execute AIC at the specified start time and intervals with.
EXTERNAL
Selects the mode of using the external start signal (external contact input) to start AIC.
For the telemeter connection specifications, if the telemeter input contact is open (telemeter malfunction),
AIC will be started using the internal clock.
OFF Selects the mode without AIC automatic start.