Product specifications

3 MEAS. SCREEN (BASIC SCREEN)
7
3 MEAS. SCREEN (BASIC SCREEN)
Note
APOA-370 uses a touch screen. Directly press keys displayed on that screen with your finger.
When pressing these keys, do not use a ballpoint pen or any other tool with a hard or sharp end. This
might cause a malfunction.
This chapter describes the MEAS. screen that is displayed immediately after the power is
turned ON.
Fig. 5 MEAS. screen
1: Icon display area
The icons showing the state of the instrument are displayed in this area.
Fig. 6 Maintenance mode icon
Note
In the case of the standard specifications, the MNT (Maintenance) signal is outputted when the
maintenance switch is ON.
Fig. 7 Mode icon
1: Icon display area
2: Current time
3: [KEY LOCK] icon (button)
4: Range display
5: Measurement result area
6: Active measurement line display
7: Function keys
Maintenance mode: This icon blinks when the maintenance switch is turned ON.
For the maintenance switch, see “ 7.1 Before Maintenance ” (page 77).
The maintenance switch is ON manually The maintenance switch is ON under external control
Mode: This icon is illuminated when gas is being sucked through any line other than
the MEAS. line.
When the gas line is switched to the MEAS. line, this icon remains illuminated
during the MEASURE time specified in the AIC sequence.