Product specifications
1 OVERVIEW
3
1.3.2 Rear panel
Fig. 3 Rear panel
Note
The measured gas is released from the exhaust outlet at a rate of 0.7 L/min.
The O
3
gas used for calibration is toxic. Be sure to connect an exhaust tube.
Name Description
1 Calibration gas inlet
The calibration gas inlet with a connector for a Teflon tube of 6 mm O.D./ 4 mm
I.D.
Make sure that the calibration gas pressure stays stable within ±500 Pa.
2 Sample inlet
The sample gas inlet with a connector for a Teflon tube of 6 mm O.D./ 4 mm
I.D.
Make sure that the sample gas pressure stays stable within ±980 Pa.
In order to prevent condensation from occurring, exercise caution to ensure
that the sample piping is not exposed to cool air.
3 Exhaust outlet
The measured gas outlet with a connector for a Teflon tube of 6 mm O.D./ 4
mm I.D.
Release the measured gas to a safe location where the back pressure stays
stable within a range of ±490 Pa.
4 RS-232C (optional)
5 Signal connection terminal block For the signals, see “ 9 EXTERNAL INPUT/OUTPUT ” (page 89) .
3
1
2
4
5