BYK-Gardner GmbH LCM plus A Instrument Description A1 General 235 017 133 DE 0302 LCM plus is a universal colorimeter for liquids suitable for manifold industrial applications. It is easy to handle and has a sturdy design. But what is really the reason why objective colour measurement of liquids is so important? The human eye is able to perceive about 100 spectral colours in approximately 100 degrees of brightness and saturation, which results in about 1 million perceptible shades.
BYK-Gardner GmbH A2 LCM plus 235 017 133 DE 0302 Instrument Design LCM plus measures transmissions of all transparent media. It operates with a halogen lamp producing standard illuminant C and uses the 2° standard observer as defined by DIN-standard 5033. A reference beam path compensates measuring values modified by driftings of lamps and temperatures during operation. Therefore, LCM plus provides an excellent long-term constancy and reproducible measuring results.
BYK-Gardner GmbH A3 LCM plus 235 017 133 DE 0302 Instrument Set-Up Unwrap the instrument carefully and check it for missing parts (delivery note). Please keep the packing box to pack the instrument safely in case it has to be transported again later. LCM plus should be placed on a level desk surface. The instrument comes with a power supply unit and adapter for the socket (100 – 240 Volt / 47 – 63 Hz) and is turned on by either pressing the OK-key or the grey menu key. Do not turn on and off too fast.
BYK-Gardner GmbH LCM plus B Operation B1 Control And Functional Elements 235 017 133 DE 0302 Figures Abb. 2 and Abb. 3 show the control and functional elements of LCM plus. Display Cuvette compartment Lamp cover Cursor keys OK-key Fig. 10: Front view Battery cover Cal- and Measure key Menu key Printer port Mains socket Bit-Box compartment Fig.
BYK-Gardner GmbH LCM plus 235 017 133 DE 0302 LCM plus is operated with 8 keys. The following table shows the keys and displays in the operating instructions.
BYK-Gardner GmbH B2 LCM plus 235 017 133 DE 0302 Initial Start-Up of the Instrument For start-up plug-in the mains plug and press the OK-key or grey menu key. The port for the power supply unit is at the rear of the instrument (page 7, Abb. 3, Items 10 and 8). When the instrument is on, the display indicates instrument type and number of EPROM-version while the self-check is running. When self-check and zero measurement are over (approx. 20 sec.
BYK-Gardner GmbH B4 LCM plus 235 017 133 DE 0302 Battery Mode LCM plus can be used with power supply unit or rechargeable batteries. For the latter mode, open the cover of the battery compartment and insert the batteries as follows: 4. 5. 6. Insert the battery with the switch to the right into the holder until only one contact remains visible. Shift the battery from the left side over the contacts, passing a slight resistance.
BYK-Gardner GmbH B5 LCM plus 235 017 133 DE 0302 Turning-Off LCM plus turns off automatically by a timer function. Manual turning-off is possible, too. Manual Turning-Off Press Cursor-Key in the measuring menu and select option „Switch Off“ with the OK-key. Then the instrument turns off. Turning-Off Via Timer On delivery, the timer of LCM plus is set to automatical turningoff time of 30 minutes.
BYK-Gardner GmbH C LCM plus 235 017 133 DE 0302 Colour Measurement Whenever the instrument is turned on, a zero measurement is made, taking the empty value (i.e. the pure transmission values without cuvette inserted) of the optical beam path and comparing it with the empty value obtained in the previous zero measurement. If the difference between both empty values is near zero, you can start measuring without calibration.
BYK-Gardner GmbH LCM plus 235 017 133 DE 0302 Calibration with 11 mm circular cuv. ok! 12:00 18.08.00 Then LCM plus prompts „Please insert sample cuvette“. Now you can insert your sample cuvette directly. As soon as you have done so, LICO starts measuring automatically. To calibrate one more path length, remove the cuvette and press the blue CAL -key again. iodine color value Please calibrate carefully to avoid mismeasurements caused by faulty calibration. 70.
BYK-Gardner GmbH C2 LCM plus 235 017 133 DE 0302 Colour Measurement The colours of all transparent liquids can be measured, using 50mm square cuvettes or 11mm round glass cuvettes, depending on their transmittances. If required, select another colour system for printer output in the configuration menu (cp. page 19). Instrument preparation 3. 4. Turn instrument and printer on. Calibrate the instrument as described in chapter. C1. Measuring procedure 1.
BYK-Gardner GmbH C3 LCM plus 235 017 133 DE 0302 Iodine Number DIN 6162 defines the Iodine colour number as mg of iodine per 100ml of potassium iodide solution. Colour matching by the Iodine number serves to judge the shade depth of clear liquids like e.g. tensides, plasticizers, resins, oils and fatty acids, whose colours resemble that of an equally layered solution of potassium iodide. For Iodine numbers less than or about 1, the Hazen number according to DIN-ISO 6271 should be preferred.
BYK-Gardner GmbH LCM plus 235 017 133 DE 0302 Measuring procedure Hazen color value 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. Measurement starts automatically. 20 _DIM ¢Conf 50° 2. After the measurement, the colour value is indicated on the display. 3. Now press the -key (DIM) until the Hazen number appears on the display. C5 Gardner Colour Number The Gardner colour number is defined in DIN ISO 4630.
BYK-Gardner GmbH LCM plus 235 017 133 DE 0302 Measuring procedure ASTM D 1500 1. 2. 3. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. After the measurement, the colour value is indicated on the display. Now press the -key (DIM) until the ASTM- number 4.5 _DIM 11° ¢Conf appears on the display.
BYK-Gardner GmbH C8 LCM plus 235 017 133 DE 0302 Configuration The configuration serves to select the colour systems for the print-out and general instrument settings like e.g. date/time, language, etc. Press cursor-key for configuration. The selection menu shown at the right opens. Move the bar to the menu options with cursor-key or . The option in reverse presentation can be selected by pressing key OK . C8.1 Switch Off LCM plus Config Printer 14:50 17.02.
BYK-Gardner GmbH LCM plus 235 017 133 DE 0302 Timer Selecting option Timer and pressing key OK opens a window to set timer functions "Aut.shut off" and "delay meas.". Set the desired value with cursor-keys and . LICO 100 Adjust timer: Turn-off Config Aut.shut off Printer delay. meas. Auto shut off time Selecting option "Aut.shut off" and pressing key 14:50 17.02.00 OK opens a window to set the turn-off time. Move the blinking Aut.
BYK-Gardner GmbH C8.3 LCM plus 235 017 133 DE 0302 Instrument check LCM plus can be checked with a certified test solution set addista ®-color in compliance with ISO 9000ff. D Technical Notes D1 Error Messages The following display messages indicate errors of operator or instrument. *** Please remove cuvette! Insert paper Error! Calibration No Signal Lamp ? Result outside the measuring range. There is a cuvette left in the cuvette compartment. Put new reel of paper into LD 500.
BYK-Gardner GmbH D2 LCM plus 235 017 133 DE 0302 Maintenande and Test Media Control Even a perfected product like LCM plus requires careful handling. Use a dry cloth or a mild household cleaner for the housing, but never solvents like naphta, acetone or the like. The optical elements should be cleaned by our maintenance staff. A certified set of test solutions is available for test media control according to ISO 9000ff with order no. 6384.
BYK-Gardner GmbH D3 LCM plus 235 017 133 DE 0302 Lamp Exchange The lamp compartment is left of the display and covered with a slotted cap on the upper side of the housing. If the lamp is broken, it should be exchanged by Dr. Lange service staff, because w e recommend a general instrument check-up after an operating period of 1-2 years. If you insist on doing it yourself, proceed as follows: R e m o v e p o w e r s u p p l y u n i t o r b a t t e r y! Remove the grey cover of the bit-box.
BYK-Gardner GmbH D5 LCM plus 235 017 133 DE 0302 Technical Data Viewing geometry: Display/operation: Reference standard: Spectral range: Wavelength accuracy: Diffuse light: Light source: Warming-up: Reproducibility: Documentation: Control: Measuring time: Program functions: Cuvettes compartment: Control elements: Data interfaces: Mains connection: Dimensions: Weight: 0°/180° (transmission) Backlit LC-display (128 x 64 Pixels) Distilled water 8 spectral wavelengths ranged from 400nm to 700nm ± 2 nm <1
BYK-Gardner GmbH LCM plus E Annex E1 Spares and Accessories Identification: Square plastic cuvettes 10mm * 50mm Square glass cuvette 10mm * 50mm Round glass cuvettes 11mm Cuvette holder for 16 round cuvettes Printer paper Exchange lamp Test solutions addista-color Rechargeable battery 1.4Ah E2 235 017 133 DE 0302 Quantity/Pcs.