Data Sheet
Product Specification Sheet APS293 - Rev. 02/2014
NC367 Wireless Pendant Receiver Unit - Momentary Contact Type
42 Central Drive • Farmingdale • NY • 11735-1202 • Phone: (631) 777-5500 • Fax: (631) 777-5599
TOLL-FREE: (800) 666-4800 • Web: AlphaCommunications.com • Email: info@alphacommunications.com
Copyright© 1998-2014 Alpha Communications® - All Rights Reserved.
Due to continuous product improvement, all colors, sizes, materials, finishes and specifications are subject to change without notice.
< Previous (APS292) Next (APS294) >
Architects' And Engineers' Specifications
The wireless receiver(s) shall be Alpha Communications® NC367
digital UHF receiver, or approved. Receiver(s) shall use the 303.875
MHz frequency for signal reception. No physical wiring connections
to the transmitter shall be required to receive a call and
the device shall include an antenna and a mounting bracket. The
receiver(s) shall have a minimum of 252 user programmable code
settings, via an easily accessible dipswitch. The receiver, upon
receipt of a signal from the WP067 wireless miniature transmitter
(or approved equal) shall momentarily actuate an internal relay.
The relay shall provide a dry 'Form C' circuit with COMMON,
NORMALLY OPEN, and NORMALLY CLOSED connections, which shall
be accessible from a clearly marked terminal strip. The relay shall
be rated for a minimum of 32 VAC of VDC @ 1 amp (NEC Class 2
Circuit).
Contractor shall supply the model SS12C (12VDC) power supply,
and shall follow manufacturer's instructions and observe all local
and national electrical and building codes.
NC367 Wireless Single Channel Receiver with
Momentary Contact
The Alpha Communications® NC367 Wireless Single Channel
Receiver with Momentary Contact is a digital UHF receiver for use
with IR300 Series (or other compatible) Patient Stations in
conjunction with the WP067 Wireless Miniature Transmitter. It
allows the user to place a 'Monitor' call by simply depressing the
button on the transmitter. This allows users to place calls
within areas of a dwelling that are not directly near nurse
signalling devices should the need arise.
The NC367 requires no wiring connections to the transmitter and
operates on the 303.875 MHz frequency. It provides a dry 'Form C'
relay circuit to integrate with the IR300 Series Patient Station. A
simple, easy-access dipswitch programming method allows for up
to 252 individual transmitter/receiver code combinations.
Features
Wireless call reception from transmitter
Digital UHF operation
252 transmitter code settings with easy
dipswitch programming
Dry 'Form C' NEC Class 2 Relay Circuit
Specifications
Dimensions:
Height: 3.80" (97mm) Width: 4.90"
(125mm) Depth: 1.30" (33mm)
Power: 12 VDC, 0.12 W Standby and 0.48W Active
Frequency: 303.875 MHz
Encoding
Technique:
Pulse width A-1 modulation at 250 bits/second;
four correct 8-bit words required for activation.
Mounting: Metal Bracket, included
Finish: Beige with light texture
Required Components
WP067 Wireless Miniature Transmitter
SS12C 12VDC Plug-In Power Supply