Specifications

Calibration Table
The calibration table contains the values measured during calibration/linearization. Individual values can
be corrected by input.
! Attention: If there is no print-out it is advisable to note down the table for the present.
Select the correction line with the ENTER key. Overwrite the line directly with the new value.
! Attention: The keyboard is directly in the number mode. If there is an input in the wrong place then
the relevant calibration value is lost. But if the calibration table is left by ESC the modications are not
valid. To get back to the old values, the instrument has to be switched off and on again.
Calibration Table Before Calibration/Linearization
ESC
M1 Enter
, ESC( )
2 %
4 %
6 %
10 %
20 %
40 %
60 %
80 %
100 %
TAR.Val.
Cal.Fact
0 %
39675
51675
63675
87675
147675
267675
387675
507675
627675
27675
30000
27675
: Abort
Delete the Table
By pressing ENTER the values, which have been read in during the calibration / linearization, are dele-
ted and the table is reset in its starting position.
! Attention: if no print-out has been made, it is adrisable to note down the table!
Print-out the Table
By pressing ENTER the calibration table is printed if a printes is available.
Page 28