Specifications
Page 26
• Sub - menu "Calibration"
Select sub-menu: factory calibration
Set cursor to calibration
Press ENTER
Set the calibration code XXXX
Select the machine.
Set cursor to the required line.
Press ENTER to reach the desired machine.
• Calibration Layout
Delete the table (in case of rst calibration only).
Reset the display to zero (rst calibration: zero reset with trimmer Z)
The calibration is executed at 80% of the measuring range.
1.
Run selected calibration value manually according to the calibration instrument asdescribed
under point "Manual operation". The display can deviate from the screen.
2.
The machine discharges. The calibration value is newly calculated. First
calibration: Run calibration value according to display. Set calibration value
with trimmer G at the calibration instrument.
CAL. FUNCTIONS
M1
Machine 1-2
Calibration
Delete calib.
calib. table
Delete table
Print cal. table
Calibr. values
Control calibrat
ESC
Linearisation
: Abort
START
ZERO
• Delete cal
The saved calibration is deleted.