Specifications
Page 15
• Automatic Test Course
Settings before the test in case of display in N/mm²
Break detection BD must be activated. The selection of the specimen must be already made, e.g.:
200,0 (cube with edge length 200 mm)
160 (cylinder with Ø 160 mm)
Course of the Test
key
Extension of piston: The peak is automatically activated and is restored
together with the break detection. The force is zeroized
and the frictional connection is detected.
Charging: The specimen is charged up to its break with a programmed
speed.
Sinking: The break detection has come into action:
- the printer prints out (option)
- the piston sinks
- the sinking time runs down (can be prematurely stopped
with STOP).
- the piston stops
-2 x STOP: the proton returns completely to its initial position.
Additional Test Courses
Change-over of parameters (option)
Setting operation (option) - e.g. the specimen is set displacement-controlled, the test is made force- or
displacement-controlled
CALC
START