Specifications

Ramp Programming for Cyclic Tests
Entrance from main menu (ramp). Here cyclic test can be programmed, but only for testing in kN (not
in N/mm²)
number of cycles determines how often the specimen is charged/discharged. With 0 the machine is
run up to the set maximal force. The force stops at this point. With STOP you can discharge again.
holding time determines how long the machine stays on the max. and min. loading points.
vel. determines how fast the specimen is charged resp. discharged
max. value determines the upper limit of the ramp
min. value determines the lower limit of the ramp
F
r holding time holding time max. value
discharge
min. value
holding time
charge
cycle 1 cycle 2
w t
F max. force
r
Remains if cycle = 0
w t
Page 14
Start
Start