Technical data
20
4. Use the user-supplied authorized Imaging
Software, such as VISIX, to activate the
Scanner and to select the desired image type
and resolution.
5. Verify that the two Scanner track indicators
illuminate green when the Scanner has been
activated, indicating that PSPs can be fed
into the corresponding Plate Guide.
Scanning and Erasing Plates
Scan and erase an intraoral PSP in one operation by
performing the following procedures.
1. Grasp a PSP by long edges between your
thumb and index finger. See Figure 9.
2. With printed side facing you, carefully and
quickly insert the PSP into the corresponding
Plate Guide slot as far as possible.
3. Immediately press the PSP all the way down
with your fingertip (from position Figure 9
to position Figure 11) until the scanner's
transport mechanism takes over and the PSP
moves down on its own.
4. Verify that the track light illuminates yellow,
indicating the Plate Guide slot is in use,
the PSP has been sensed and the Scanner is
transporting the PSP.
Figure 9. PSP Pick-up.
Figure 10
Feeding Intraoral Plate.
Figure 11.
Fully Inserted Plate.
IMPORTANT: Transfer the PSP from the PSP Table Mat to the Plate
Guide slot quickly. Always keep the sensitive side away
from any light source to minimize image loss.
Activate Scanner
Activate the ScanX by performing the following procedures.
1. Make sure the ScanX and computer components are correctly connected
as shown in Figure 4.
2. Switch the scanner from Standby to ON by pressing the membrane READY
switch ( ) located on the Keypad Control Panel on the top of the scan-
ner.
3. Verify that both the green and blue LED indicators above the READY and
ERASER switches, respectively, illuminate. (Default mode has eraser ON.)
Note: Change in motor noise pitch levels is normal when switching to different
resolution setting selections.
INTRAORAL IMAGING PROCEDURES