Technical data

20
MAINTENANCE
Maintenance Procedures
The ScanX is designed for many years of trouble-free operation. Maintenance as
described herein is minimal.
IMPORTANT: Do not spray solvents or liquid directly
on the scanner.
Cleaning the ScanX
Turn off the ScanX disconnect the line cord from the Mains wall outlet and dis-
connect the computer connection cable from the ScanX before cleaning. Wipe
the outside surfaces with a soft paper towel dampened with a disinfectant solu-
tion or non-abrasive household cleaner. Be careful not to allow solvents TO RUN
OR DRIP into the ScanX. This could cause damage to the ScanX. Allow to air dry
before plugging in or turning back on.
Cleaning the Plate Transport
Over time, small debris and dust can accumulate in the plate transport mecha-
nism causing a loss in image quality and possible damage to the PSPs.
To ensure optimal performance of the ScanX, the plate transport should be
cleaned once per week using a new ScanX Cleaning Sheet each time. Sample
sheets are included with the ScanX and additional sheets can be purchased from
your dealer. Refer to ACCESSORIES page 23.
Disinfecting the ScanX
No disinfection is required for the ScanX.
POWERING DOWN THE SYSTEM
The ScanX is designed to be left on continuously during the active day. At the
end of the day, or whenever desired, power down the system as follows:
1. Switch the ScanX from ON to Standby bypressing and holding the
membrane READY switch ( ) on the Keypad Control Panel for
approximately two seconds, until the green LED above the READY
switch extinguishes.
2. Turn OFF the main power to the ScanX by placing the rocker switch
on the Built-In Control/Connector panel to the OFF (O) position.
IMPORTANT: Never power down the system during a scanning
session.