Technical data

12
TECHNICAL DATA
Electrical Requirements:
Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz
Supply Current: 2A Maximum
Line Cord: Country Specific Hospital Grade line cord,
(See Note).
Physical Properties:
Operating Dimensions:
In Opened Soft Case: D 17 x W 34 x H 19 3/8 inches
Out of Soft Case: D 17 x W 15 5/8 x H 13 3/4 inches
Storage Dimensions: D 17 x W 17 x H 14 1/2 inches
Weight: 36 lbs.
Environmental Conditions:
Unit in Operation
Temperature: 0 to 104°F (-18 to 40°C)
Humidity: 5% to 95% RH
Heat emission: <90W
Storage and Transport
Temperature: -21 to 130°F (-29 to 54°C)
Humidity: 5% - 95% (Non-condensing)
Note: Resolution of the ScanX is dependent on operating mode and specific imaging plate type
used.
Resolution (LP/mm)
Vertical 5.0
Horizontal 5.0
Compliance Data:
Laser Class: Class I Laser Product (21 CFR 1040.10)
Class 1 Laser Product (IEC 60825-1)
Laser Product Report
Accession Number: 0212282-00
Installation Category: II
Pollution Degree: 2
User Replacement Items: Line Cord, PSPs
and Computer Connector Cable
IEC601-1 Classification:
Class 1, Type B, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic
mixture(s). Protection against ingress of liquids -Ordinary
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can
occur. Do not use the equipment in close conjunction with sensitive devices, or
devices creating high electromagnetic disturbances.
Note: The ScanX is equipped with a line cord matching Mains outlet of country of intended use.
USA and some international power line cords are available as required.