Technical data
12
TECHNICAL DATA
Electrical Requirements:
Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz
Supply Current: 1.2 A Maximum
Line Cord: North American style 10 foot long
Hospital Grade power cord, P/N 73096.
Power Supply: 24 Volt Power Supply, P/N B7095 provided
Physical Properties:
Environmental Conditions:
Unit in Operation
Temperature: 50°F to 105°F (10°C to 40°C)
Humidity: 5% to 95% RH
Heat emission <90W
Storage and Transport
Temperature: -21°F to 130°F (-29°C to 54°C)
Humidity: 5% - 95% (Non-condensing)
Note: Resolution of the ScanX is dependent on operating mode and specific imaging plate type
used.
Resolution Range: P/N B7110V P/N B7310V
Horizontal: 10 to 14 (lp/mm) 2.5 to 14 (lp/mm)
Vertical: 10 to 18+ (lp/mm) 2.5 to 18+ (lp/mm)
Compliance Data:
Laser Class: Class I Laser Product (21 CFR 1040.10)
Class 1 Laser Product (IEC 60825-1)
Laser Product Report
Accession Number: 0212282-00
Installation Category: I
Pollution Degree: 2
User Replacement Items: Power Supply, Line Cord, Plate Guides, PSPs
and Computer Connector Cable
IEC 61010-1 Classification:
Class 1, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic
mixture(s). Protection against ingress of liquids -Ordinary
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can
occur. Do not use the equipment in close conjunction with sensitive devices, or
devices creating high electromagnetic disturbances.
ScanX Part Number B7110V B7310V
Depth (inches) 15.13 15.4
Width (inches) 15.0 15.0
Height (inches) 15.5 27.8
Weight (pounds) 43 50