Operator`s manual

5
SAFETY INSTRUCTIONS
DANGER: Opening the ScanX by removing any covers or components makes the equipment
into a Class III b Laser Product. [Class 3B Laser Product (IEC)].
WARNINGS -
The ScanX contains a laser. The ScanX is a Class I [Class 1 (IEC)] Laser Product.
Use of controls or adjustments or performance of procedures other than those specified herein may
result in hazardous radiation exposure. The laser is on only during an active scan.
Only a trained technician from an authorized ALLPRO Imaging dealer
should remove a cover from the ScanX.
Direct eye contact with the output beam from the laser may cause serious damage and possible
blindness.
Do not open the ScanX to maintain it.
The ScanX contains no user serviceable parts. If there is a service problem, contact your authorized
ALLPRO Imaging dealer.
Use of Accessory Equipment.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this
equipment may lead to a reduced level of safety of the resulting system.
Consideration relating to the choice shall include:
- use of the accessory in the PATIENT VICINITY
- evidence that the safety certification of the ACCESSORY has been performed in accordance
to the appropriate IEC 601-1 and/or IEC
601-1-1 harmonized national standard.
Use of ACCESSORIES or cables other than those specified or provided by ALLPRO Imaging may
result in increased EMISSIONS or decreased IMMUNITY of the EQUIPMENT.
Operate ScanX in dry environment.
To prevent fire or electrical shock, do not expose this appliance to rain or moisture.
Use only grounded electrical connections.
Use only the power cord provided with the unit.
Do not use damaged Phosphor Storage Plates (PSPs).
Damaged PSPs may not provide reliable diagnostic images.
Disposal of PSPs.
Consult with your federal, national, state and local government, for rules and regulations on
disposal of PSPs.
CAUTIONS -
Completely clean and erase PSPs before taking an X-ray exposure.
See the PLATE PREPARATION section of this manual.
Minimize exposing an X-ray exposed PSP to light.
Transfer the PSP to the Inlet slot within the Plate Protector and/or quickly and with minimal
light exposure.
Use care in handling PSPs - Avoid fingerprints and scratching
Refer to the instructions provided with the PSP package for further information on handling.
Use of Other Manufacturer’s Imaging Plates
Do not put PSPs designed for drum-type or other scanners in the ScanX. The hooks and/or frames
on the ends or around these PSPs, or PSPs of different thickness (especially thicker ones) will damage
the ScanX. PSPs from another manufacturer may be used as long as the specifications are the same
as those for the ScanX PSPs.