Note: 1. Cassettes must not contain intensifying screens when using PSPs.
2. PSPs may be exposed in cassettes and scanned without Plate
Protectors, however, greater care must be exercised to not expose the
PSP to light before scanning and to prevent the PSP from being
scratched or soiled.
Take an X-ray Image. Put an image on the PSP by performing the following procedure.
1. Load cassette containing an erased PSP in a Plate Protector into the exposure device as
previously done with film.
2. Take the exposure using either the same settings used for film if previously “green” film
was used or with about half the exposure if “blue” film was used with intensifying
screens.
3. Bring the closed cassette housing the X-ray exposed PSP containing the latent image to the
ScanX. The PSP is now ready to be scanned to read the latent image from the PSP.
PLATE PREPARATION
Black Side or Back of PSP
White or Sensitive Side
or Front of PSP
Prior to performing Imaging procedures provided on the following pages, the user must be familiar
with the care and handling of the PSP in order to ensure successful image scanning. Figure 5
shows the configuration of a typical 10 x 12 inch PSP.
CAUTION: PSPs must always be erased prior to use.
Erase the PSP. Each PSP should be used (i.e. X-ray exposed and scanned) within 24 hours
of erasure since natural radiation will add noise to the PSP. ALLPRO Imaging recommends the
use of the ALLPRO ScanX Eraser (p/N B2840) for PSPs measuring a maximum of 14 X17 inches
with an erasing time of approximately 15 seconds. Equivalent erasing is achieved with 750,000
lux-seconds of erasing energy.
CAUTION: Use a Plate Protector and exercise care in handling the plate so
as not to scratch the sensitive surface or nick the edges.
Plate Protector. Always exercise care in handling the PSP so as not to scratch or soil the
sensitive surface or nick the edges. Make sure to use a correct size Plate Protector with the PSP.
X-ray Cassette. Place the PSP covered with a Plate Protector into the appropriate X-ray cassette
with the sensitive (front) side of the PSP towards the Tube-side of the cassette and close cassette.
FFiigguurree 55.. PPSSPP oorr IImmaaggiinngg PPllaattee CCoonnffiigguurraattiioonn
IMAGING PROCEDURES