Operator`s manual

TO
MAINS
OUTLET
11
SYSTEM SETUP
SETUP PROCEDURE
Perform the following procedure when setting up the ScanX for operation for the first time.
1. Select a location that meets the SITE SELECTION guidelines.
2. Install the ScanX into countertop per instructions provided with the In-Counter Installation
Kit, P/N B2988
3. Set up the computer according to the manufacturer's recommendations.
4. Install ALLPRO Imaging Authorized Image Management Software on the computer.
5. Refer to figure 4 and connect the line cord between the ScanX power IEC connector
located on the Built-In Control/Connector Panel and the Mains outlet.
6. Connect the high speed USB cable between the USB Type B connector located on the Built-
In Control/Connector Panel and the USB Type A connector located on the computer. Always
make sure to use the same USB port whenever reconnection of the USB cable is necessary
.
7. Apply the main power to the ScanX by placing the rocker switch on the Built-In
Control/Connector panel to the ON (I) position.
8. Switch the scanner from standby to ON by pressing the membrane switch ( ) on the
top of the scanner. Verify that the green light below the switch illuminates.
9. When prompted by Windows, insert the supplied ScanX Drivers and Utilities Disk which
contains the USB drivers and install per the Windows Setup Wizard instructions.
Main Power
Switch (1/0)
(See Note)
IEC
Connector
Power
Cord
(See Note)
1A Circuit
Breakers
USB Connector
Type B
Scanner ON
Switch
Scanner
ON LED
Scanner Track
Indicator LEDs
FFiigguurree 44.. SSccaannXX IInnssttaallllaattiioonn aanndd CCoonnttrrooll//IInnddiiccaattoorrss LLooccaattiioonn DDiiaaggrraamm
USB TO
COMPUTER
Note:
The Power Cord is the
Mains disconnect device.