Technical data

12
TECHNICAL DATA
Electrical Requirements:
Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz
Supply Current: 1.2 A Maximum
Line Cord: North American style 10 foot long
Hospital Grade line cord, P/N 73096.
Power Supply: 24 Volt Power Supply, P/N B7095 provided
Physical Properties:
Dimensions
Depth: 15.4 in.
Width: 15.0 in.
Height: 27.8 in.
Weight: 50 lbs.
Environmental Conditions:
Unit in Operation
Temperature: 50°F to 105°F (10°C to 40°C)
Humidity: 5% to 95% RH
Heat emission: <90W
Storage and Transport
Temperature: -21°F to 130°F (-29°C to 54°C)
Humidity: 5% - 95% (Non-condensing)
Note: Resolution of the ScanX is dependent on operating mode and specific
imaging plate type used.
Resolution Range: B7310 B7310-HR
Horizontal: 3.8 to 5+ (lp/mm) 3.8 to 14 (lp/mm)
Vertical: 3.8 to 5+ (lp/mm) 3.8 to 18+ (lp/mm)
Compliance Data:
Laser Class: Class I Laser Product (21 CFR 1040.10)
Class 1 Laser Product (IEC 60825-1)
Laser Product Report
Accession Number: 0212282-00
Installation Category: I
Pollution Degree: 2
User Replacement Items: Power Supply, Line Cord, PSPs
and Computer Connector Cable
IEC 61010-1 Classification:
Class 1, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic
mixture(s). Protection against ingress of liquids -Ordinary
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can
occur. Do not use the equipment in close conjunction with sensitive devices, or
devices creating high electromagnetic disturbances.