Specifications
Basic parameters and commands
Pearleye Technical Manual V2.0.0
41
The camera is equipped with an internal temperature sensor. The temperature
can be displayed using the command T=2. The temperature is output as 16 bits
value. The 4 most significant bits contain information about the temperature
sensing state. The remaining 12 bits contain two’s complement of 1/16 degrees
Celsius.
Software version and correction data
information (V)
By means of the command V=1 information about the correction data installed
in the non-volatile memory is displayed. The current firmware version and also
the serial number of the camera are output.
Note
For information about the starting procedure see Warm-up
period on page 10.
GigE feature name (CameraSpecialFeatures) Feature Visibility Description
SensorTemperatureState Expert Camera sensor temperature
state.
0 = The sensor temperature is
OK.
1 = The sensor temperature is
outside the optimum range.
QuerySensorTemperatureState Expert Query camera sensor tempera-
ture state.
(T = 1 command)
ShutterTemperatureValue Beginner Shutter temperature value in
degree Celsius.
QueryShutterTemperatureValue Beginner Query camera shutter tempera-
ture state.
(T = 2 command)
Table 15: GigE feature: Sensor temperature state
Note
This function is not available as GigE Vision feature.