Specifications
Basic parameters and commands
Pearleye Technical Manual V2.0.0
40
Electromechanical shutter (I)
The Pearleye P-007 LWIR / Pearleye P-030 LWIR models are equipped with an
electromechanical shutter that can be controlled with the command I.
Use I=1 to close and I=0 to open the shutter.
Temperature warning (T)
The command T=1 displays the content of the temperature warning register. The
lowest order bit has the same function as LED L3 placed at the backside of the
camera housing. In case the value 1 is output, the temperature of the microbo-
lometer sensor is outside the optimum range. After switching-on the cold cam-
era this may take place for a time because the stabilization of the sensor tem-
perature takes some minutes until the set value becomes achieved. In case that
the temperature warning remains active for a longer period, the environment
temperature of the camera should be checked.
GigE feature name (CameraSpecialFeatures) Feature Visibility Description
AutoCalibrationInterval Expert Setup the automatic calibra-
tion interval.
0=Calibrate one-time,
1...65535=Calibrate every
k*256 frames.
(k=<value> command)
Table 13: GigE feature: AutoCalibrationInterval
Note
The shutter is open by default (power off).
GigE feature name (CameraSpecialFeatures) Feature Visibility Description
MechanicalShutter Beginner Set the electromechanical
shutter state manually.
(I=<value> command)
Table 14: GigE feature: Electromechanical shutter